Measurement of Whole Blood Tacrolimus Concentrations by LC-MS/MS and Immunoassay Methods: Influence of
Adekunle Alabi1, Mengyuan Ge1, Jeremiah D Momper1
1Department of Pathology, UC San Diego Health, La Jolla, CA, United States.
Background:
Therapeutic drug monitoring of the immunosuppressant tacrolimus is commonly performed by immunoassay or LC-MS/MS. Measurement biases between these methodologies have been characterized for immediate-release tacrolimus (IR-tac; Prograf) but have not been performed for extended-release formulations such as Envarsus. These discrepancies can impact patient care, as appropriate dosing is required to maintain therapeutic concentrations and immunosuppression.
Methods:
Validation of a whole-blood LC-MS/MS method for the simultaneous quantification of tacrolimus and its major metabolite, desmethyl tacrolimus, was performed using traceable calibrators (tacrolimus, ERM-DA110a) and quality control (QC) material for tacrolimus and standard material for desmethyl tacrolimus. Tacrolimus concentrations were determined by LC-MS/MS and the ARCHITECT immunoassay in patients receiving either IR-tac or Envarsus for clinical care.
Results:
External calibration curves for both tacrolimus and desmethyl tacrolimus were linear (R2 > 0.995), and the analytical measurement range (AMR) for tacrolimus spanned from 1.1 to 31.6 ng/mL. Calibrator/QC biases were within 15% of their spiked concentrations throughout the AMR, and within-run imprecision was <10%, except at the lower limit of quantification (n = 25). Between-run imprecision for low, mid, and high QC levels was ≤11% over a 2-week period (n = 5 days). Comparative biases between immunoassay and LC-MS/MS were significantly lower (P = 0.0074) for patients receiving Envarsus (n = 20 specimens) relative to patients receiving IR-tac (n = 32 specimens).
Conclusions:
Biases between immunoassay and LC-MS/MS tacrolimus measurements in patients receiving immediate-release vs extended-release formulations indicate that their distinct pharmacokinetic profiles impact measurement accuracy. These assay biases should be considered when interpreting tacrolimus concentration measurements.
Related Concept Videos
Drug Concentration Versus Time Correlation
Two pivotal parameters are the minimum effective concentration (MEC) and the minimum toxic concentration (MTC). The MEC is the lowest drug...
Drug Concentrations: Measurements
Plasma —...
Measurement of Bioavailability: Pharmacokinetic Methods
Measurement of Bioavailability: Pharmacodynamic Methods
Therapeutic Drug Monitoring: Affecting Factors
Therapeutic Drug Monitoring: Drug Analysis Methods