Reverse Torque and Wear of Internal Connections of Metal-Based Abutments Under Cyclic Loading
Daher Antonio Queiroz1, Mariana Itaborai Moreira Freitas2, Kaden Thomas Ockey1
1Department of Restorative Dentistry & Prosthodontics, The University of Texas Health Science Center at Houston (UTHealth) School of Dentistry, Houston, Texas, USA.
Objective:
Evaluate the reverse torque values and wear of internal connections caused by metal-based zirconia abutments under cyclic loading.
Material And Methods:
Thirty implants were divided into three groups (n = 10), based on connection systems: Interference-fit cone-screw (Cortical Master Flash MI Fit-MTA and Cortical Master Flash MI RP-MTB), and Internal hexagon (Cortical Titanium Master Connect AR-IH). Metal-based zirconia abutments were tightened into the implant, and zirconia crowns were bonded using resin cement. All specimens underwent cyclic loading at 300 N and 9 Hz for 1,000,000 cycles. Reverse torques were assessed postloading, and wear was analyzed using scanning electron microscopy (SEM). Statistical analysis employed one-way ANOVA for reverse torque and wear values, with significant differences assessed via the Tukey test (α = 0.05).
Results:
Significant differences in reverse torque values were found among groups: MTA (24.9 N), MTB (24.1 N), and IH (14.4 N). The IH group showed higher mean wear values on the tension side (121.0) and compression side (112.6), with a wear difference of 8.4.
Conclusion:
The interference-fit cone-screw abutments demonstrated significantly reduced torque loss and morphology after cyclic loading, reflecting enhanced performance.
Clinical Significance:
Although the stability of the metal base zirconia abutments to the implants was accurate, clinicians should be cautious about torque loss and more wear when using internal hexagon connections.
More Related Videos
03:07Scanning Electron Microscopic Evaluation of Surface Defects of Remover Retreatment File After Single and Multiple Uses
Published on: October 11, 2024
07:09In situ Compressive Loading and Correlative Noninvasive Imaging of the Bone-periodontal Ligament-tooth Fibrous Joint
Published on: March 7, 2014
Related Concept Videos
Residual Stresses in Circular Shafts
Stresses in a Shaft
Applying equilibrium conditions to the QR segment establishes that the internal shearing forces within the...
Screw: Problem Solving
To evaluate the...
Stress Concentrations in Circular Shafts
Deformation in a Circular Shaft
Thin-Walled Hollow Shafts