Rapid Point-of-Care Assay of Enoxaparin Anticoagulant Efficacy in Whole Blood
Published on: October 12, 2012
13.7K
Low-Range Heparin and Protamine Detection: A Single-Center Prospective Diagnostic Study.
Michael Vandenheuvel1, Laura Vierstraete2, Filip De Somer3
1Department of Anaesthesia, University Hospital Ghent, Ghent, Belgium.
Journal of Cardiothoracic and Vascular Anesthesia
|January 31, 2025
Summary
The Sonoclot slope-45 accurately detects low heparin levels, even with protamine present. This new parameter offers a sensitive and reliable method for monitoring heparin activity post-surgery.
Area of Science:
- Anesthesiology and Critical Care Medicine
- Hematology
- Biomedical Engineering
Background:
- Accurate detection of low-range heparin activity is crucial for managing heparin reversal and rebound, particularly after cardiopulmonary bypass.
- Existing methods like anti-Xa activity (aXa), activated partial thromboplastin time (aPTT), kaolin-activated clotting time (kACT), and ROTEM viscoelastic tests have limitations in sensitivity, point-of-care availability, or cost.
- The Sonoclot viscoelastic test offers a potential alternative for monitoring heparin activity.
Purpose of the Study:
- To evaluate the performance of the Sonoclot slope-45 parameter for detecting low-dose heparin activity.
- To assess the impact of protamine, alone and in combination with heparin, on various coagulation parameters, including the Sonoclot slope-45.
- To compare the sensitivity of the Sonoclot slope-45 against established coagulation tests.
Main Methods:
- A prospective in vitro study was conducted using blood samples from 10 healthy volunteers.
- Blood samples were treated ex vivo with incremental low doses of heparin, protamine, or a combination of both.
- The effects on anti-Xa activity (aXa), activated partial thromboplastin time (aPTT), iSTAT kaolin-activated clotting time (kACT), ROTEM clotting time ratio, and the Sonoclot slope-45 were analyzed using receiver operating curve analysis.
Main Results:
- The Sonoclot slope-45, along with aXa and aPTT, reliably detected low heparin activity in spiked blood samples.
- The iSTAT kACT was found to be insensitive to low heparin concentrations.
- Protamine, whether administered alone or with heparin, did not significantly affect any of the tested coagulation parameters, including the Sonoclot slope-45.
Conclusions:
- The Sonoclot slope-45 demonstrates high sensitivity for detecting low-dose heparin activity.
- This sensitivity is maintained even in the presence of protamine, suggesting its utility in assessing heparin reversal.
- The Sonoclot slope-45 was found to be insensitive to protamine alone, further clarifying its specific role in heparin monitoring.