Safety and efficacy of a temperature-controlled ablation system for ventricular tachycardia: Results from the TRAC-VT
Josef Kautzner1, Javier Moreno2, Claudio Tondo3,4
1Institute for Clinical and Experimental Medicine, Prague, Czechia. josef.kautzner@ikem.cz.
Background:
Catheter ablation using radiofrequency (RF) energy is an established treatment for ventricular tachycardia (VT). Tissue temperature is a key determinant of successful lesion creation, and yet, it is difficult to measure during conventional RF ablation because of the cooling effect of high-flow rate saline irrigation. The TRAC-VT study evaluated the safety and efficacy of a novel irrigated RF ablation system modulating power based on real-time tissue temperature.
Methods:
Patients with sustained monomorphic VT and structural heart disease (SHD) were enrolled. Catheter ablation was performed in temperature-control mode (irrigation 8 ml/min, temperature set-points 55 or 60 °C, and power output ≤ 50 W), with RF applications for ≤ 45 s. The primary safety endpoint was a composite of cardiovascular-specific serious procedure-related adverse events within 30 days post-ablation. The primary effectiveness endpoint was acute success (i.e., non-inducibility of all clinically relevant VTs).
Results:
Thirty-eight patients were enrolled with monomorphic VT (age 68 ± 12 years and 84% male), with an average of 1.7 ± 1.2 VTs targeted per patient. In total, 41 ± 23 RF applications per patient were delivered. Acute procedural success was 100% (95% CI, 91-100%). No primary safety endpoints were observed. Six-month follow-up was completed in 92% of patients with 81% (95% CI, 65-91%) freedom from sustained or treated VT. A repeat ablation was performed in three patients.
Conclusions:
Ablation of VT in SHD, using a temperature-controlled irrigated RF catheter, was safe and effective with a low rate of VT recurrence at 6 months.
More Related Videos
06:25Cooling or Warming the Esophagus to Reduce Esophageal Injury During Left Atrial Ablation in the Treatment of Atrial Fibrillation
Published on: March 15, 2020
04:58Reduced Procedure Time and Variability with Active Esophageal Cooling During Radiofrequency Ablation for Atrial Fibrillation
Published on: August 25, 2022
Related Concept Videos
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Autoregulation of Blood Flow
Chemical Signaling in Autoregulation
Chemical signaling operates at the precapillary sphincter level, inciting either contraction or relaxation....