Repeatability of tear film lipid layer interferometry measurements: A randomized, crossover study
Michael T M Wang1, Stuti L Misra1, Akilesh Gokul1
1Department of Ophthalmology, Aotearoa New Zealand National Eye Centre, The University of Auckland, Auckland, New Zealand.
The Keratograph 5M and LipiView II provide reliable tear film lipid layer measurements. Both qualitative grading and quantitative thickness assessments show good repeatability and reproducibility for clinical use.
Area of Science:
- Ophthalmology
- Ocular Surface Disease
- Biomedical Optics
Background:
- The tear film's lipid layer is crucial for ocular surface health.
- Accurate measurement of the lipid layer is important for diagnosing and managing dry eye disease.
- Existing methods for assessing the lipid layer include qualitative grading and quantitative thickness measurements.
Purpose of the Study:
- To evaluate the intrasession repeatability and diurnal/day-to-day reproducibility of tear film lipid layer interferometry.
- To compare qualitative grading (Keratograph 5M) and quantitative thickness measurements (LipiView II).
- To assess these measurements in a general community population without pre-existing dry eye or blepharitis.
Main Methods:
- A randomized, crossover study involving 40 healthy community residents.
- Measurements were taken during three sessions: two on the same day (6 hours apart) and one on a separate day.
- Tear film lipid layer grading and thickness were measured three times per session with 10-minute intervals.
Main Results:
- No significant differences were found in lipid layer grading or thickness across all comparisons (p≥0.19).
- High correlations (p<0.001) and good reproducibility (Cohen's κ: 0.670–0.867; ICC: >0.80 for intrasession, 0.60–0.80 for intervisit) were observed.
- Quantitative thickness measurements showed low Bland-Altman biases (around 3 nm) and test-retest repeatability of 4.5–11.7 nm.
Conclusions:
- Both qualitative and quantitative tear film lipid layer interferometry methods demonstrate favorable repeatability and reproducibility.
- These findings support the use of either method in clinical and research settings for reliable lipid layer assessment.
- The study validates these interferometry techniques as dependable tools for evaluating the tear film lipid layer.