Orthopaedic implant cobaltism from spinal instrumentation
Jack Donovan Sedwick1, Stephen Tower2, Robert Bridges3
1University of Washington School of Medicine, Seattle, Washington, USA jsedwick@uw.edu.
BMJ Case Reports
|February 3, 2025
Summary
Orthopaedic implant cobaltism (OIC) from spinal implants is reported for the first time. This case shows OIC is reversible after removing cobalt-chromium components, improving patient symptoms and blood-cobalt levels.
Area of Science:
- Orthopaedic Surgery
- Toxicology
- Biomaterials Science
Background:
- Cobaltism is a known complication of cobalt-chromium (CoCr) implants, typically hip and knee arthroprostheses.
- Spinal implants made of CoCr alloys have not been previously associated with cobaltism.
- Orthopaedic implant cobaltism (OIC) severity correlates with cobalt levels in blood and urine and implant wear/corrosion.
Purpose of the Study:
- To report the first case of OIC resulting from failed spinal instrumentation.
- To highlight the potential for CoCr spinal components to cause systemic cobalt toxicity.
- To demonstrate the reversibility of OIC following revision surgery.
Main Methods:
- Case report of a patient with scoliosis instrumentation failure.
- Analysis of peri-implant fluid and tissue for cobalt levels and metallosis.
- Clinical assessment of neurological and visual symptoms.
- Revision surgery to remove CoCr implants and replace them with non-cobalt containing components.
Main Results:
- The patient presented with tremors, cognitive decline, insomnia, and visual changes.
- High cobalt levels (>4800 mcg/L) in peri-implant fluid and evidence of metallosis confirmed OIC.
- Post-revision surgery, the patient's symptoms resolved, and blood-cobalt normalized.
Conclusions:
- Cobalt-chromium spinal implants can cause OIC, a spectrum illness with potentially severe neurological and visual symptoms.
- This case demonstrates that OIC secondary to spinal implants is reversible upon removal of the offending components.
- This is the first reported instance of OIC originating from spinal implants.