Performance of Three Continuous Glucose Monitoring Systems in Adults With Type 1 Diabetes
Manuel Eichenlaub1, Delia Waldenmaier1, Stephanie Wehrstedt1
1Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm, Ulm, Germany.
Background:
The performance of continuous glucose monitoring (CGM) systems is difficult to compare due to different study designs and a lack of head-to-head studies. This study evaluated the performance of FreeStyle Libre 3 (FL3), Dexcom G7 (DG7), and Medtronic Simplera (MSP) against different comparator methods and during clinically relevant glycemic scenarios.
Method:
Twenty-four adult participants with type 1 diabetes mellitus wore one sensor of each CGM system in parallel for up to 15 days. Sensors of DG7 and MSP were exchanged on days 5 and 8, respectively. Three 7-hour sessions with 15-minute comparator blood glucose-level measurements using YSI 2300 (YSI, venous), Cobas Integra (INT, venous), and Contour Next (CNX, capillary) were conducted on days 2, 5, and 15. Simultaneously, glucose-level excursions with transient hyperglycemia and hypoglycemia were induced according to a recently published testing procedure. The accuracy was evaluated using various metrics, including mean absolute relative differences (MARDs).
Results:
Compared with YSI data, the MARDs of FL3, DG7, and MSP were 11.6%, 12.0%, and 11.6%, respectively. Relative to the INT data, the corresponding MARDs were 9.5%, 9.9%, and 13.9%, respectively, and compared with CNX data, MARDs were 9.7%, 10.1%, and 16.6%, respectively. Both FL3 and DG7 showed better accuracy in the normoglycemic and hyperglycemic range, while MSP performed better in the hypoglycemic range.
Conclusions:
Performance results of all CGM systems varied depending on the comparator method. However, across comparators FL3 and DG7 tended to be more accurate compared with MSP. All CGM systems showed a lower accuracy compared with previous studies, emphasizing the need for comprehensive study design guidelines.
Related Concept Videos
Diabetes Mellitus: Type 2 and Gestational
Diabetes: Symptoms, Diagnosis, and Complications
Diabetes Mellitus: Overview and Type I Subtype
Type 1 diabetes is an autoimmune disease in which the immune system mistakenly attacks and destroys the insulin-producing beta cells in the pancreas. As a result, the body is unable to produce sufficient insulin, and individuals with...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Diabetes: Management and Pharmacotherapy
Insulin remains the cornerstone of treatment for most patients with type 1 and many...
Hypoglycemia and Glucagon