Absorbed dose differences between twin fetuses for pregnancy patients in CT examinations
Shuiyin Qu1,2, Haoran Jia1, Habib Zaidi3,4,5,6
1Institute of Radiation Medicine, Fudan University, Shanghai, China.
Background:
Estimation of the radiation dose to the fetus is essential for the assessment of radiation risks and benefits to pregnant patients undergoing radiological examinations. During the past decade, the global twinning rate has soared resulting from embryo assistance and increased delivery age. However, to the best of our knowledge, radiation dosimetry in multiple pregnancies from radiological imaging has never been reported before.
Purpose:
The purpose of this study is to develop personalized computational models for twin fetuses based on clinical CT images of real pregnant patients and to estimate personalized radiation doses for twin fetuses from abdominal/pelvic CT examinations.
Methods:
Personalized computational phantoms representing pregnant females with twins at the second and third trimesters were constructed based on CT images of two pregnant patients. Monte Carlo calculations were performed using the MCNP transport code and three validated CT scanners to estimate the radiation dose of twin fetuses during abdominal and pelvic CT examinations.
Results:
The absorbed fetal organ dose was calculated and compared between twins. For the same patient, the absolute difference in fetal organ dose between twins varies between 0.63% and 39.64% with an average value of 12.85%. The estimated total-body dose differences for twin fetuses were 11.55% and 7.51%, respectively, for pregnant patients at 22 and 30 weeks gestational age.
Conclusion:
The variations of body weight and organ mass affect the absorbed dose of twin fetuses. Personalized computational models provide more accurate fetal radiation dosimetry estimates for pregnant patients with twins. This work also contributes to a better understanding of model-induced uncertainties in external radiation dosimetry for the developing fetus.
More Related Videos
09:52Accurate and Simple Evaluation of Vascular Anastomoses in Monochorionic Placenta using Colored Dye
Published on: September 5, 2011
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Related Concept Videos
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...