Aim Of The Study:
The objective of the present study was to establish local reference data for radiation exposure in children undergoing port implantation.
Methods:
A retrospective analysis of radiation exposure was conducted, with the primary endpoint being the third quartile of the dose area product (DAP) measured in cGy · cm2. The study included all pediatric port implantations performed at our institution between 2018 and 2023. Secondary endpoints comprised patient age, underlying diagnoses, radiation duration, and duration of surgery. All procedures were performed under fluoroscopy using a mobile C-arm with a flat panel detector (Ziehm Vision, Leu AG). This device automatically adjusts power output based on tissue density. The standard pre-installed program, "Children and Heart", delivers 12 pulses per second. At our institution, pediatric surgeons are responsible for intraoperative fluoroscopy. Mandatory annual radiation protection training is provided for all surgeons utilizing fluoroscopic equipment.
Main Results:
A total of 157 port implantations in 143 patients were analyzed. Of these, 57 % (n = 82) were male and 59 % (n = 93) of the procedures were conducted on patients younger than seven years. The most common indications for port implantation were leukemia (41 %, n = 59) and brain tumors (16 %, n = 23). The value for the third quartile of the DAP was 24.6 cGy · cm2, with calculated effective doses ranging from 0.009 mSv to 0.023, depending on patient weight. The mean radiation duration was 31 s. The radiation exposure was found to be positively correlated with patient's weight.
Conclusion:
The literature on radiation exposure in pediatric port implantation is limited. The present analysis provides a benchmark for children and adolescents, useful for quality control and the development of protocols that aim to reduce radiation exposure in children.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
08:34Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
Related Concept Videos
Biological Effects of Radiation
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
X-ray Imaging
Mutations
Chromosomal Alterations Are Large-Scale Mutations
While point mutations are changes in a single nucleotide in...
Radiological Investigation I: X-ray and CT