High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
Positron Range Correction Helps Enhance the Image Quality of Cardiac 82Rb PET/CT
Martin Lyngby Lassen1,2, Hunor Kertész3, Ivo Rausch4
1Department of Clinical Physiology and Nuclear Medicine, University Hospital Copenhagen-Rigshospitalet, Copenhagen, Denmark; martin.lyngby.lassen@regionh.dk.
None:
The image quality and quantitative accuracy of 82Rb myocardial perfusion imaging (MPI) using PET is challenged by the extensive positron range (PR) effects, with the PR of 82Rb being about 7 mm in soft tissues. This study explored the feasibility of applying postacquisition PR correction (PRC) to routine 82Rb PET/CT MPI acquisitions and assessed its impact on diagnostic accuracy and image quality. Methods: We implemented a PRC method adjusted to 82Rb into a vendor-provided reconstruction toolbox, using tissue-specific corrections for soft tissue, bone, and air/lungs. The PRC was evaluated in 2 cohorts: the first comprised 25 healthy volunteers who underwent repeated 82Rb MPI within 2 wk, and the second included 66 patients with known or suspected coronary artery disease. We measured the signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) for the volunteer cohort. In the patient cohort, the impact of PRC was evaluated as changes in the area under the receiver operating characteristic curve (AUC), using fractional flow reserve as the gold standard (values < 80% were considered significantly reduced). We calculated AUCs for stress and ischemic total perfusion deficits. Results: In the volunteer cohort, PRC-based reconstructions (standard reconstruction [STD] + PRC) demonstrated significantly improved SNR and CNR compared with STD, with median increases of 22% and 47% for SNR and CNR, respectively (P < 0.05). For the patient cohort, comparable AUCs were reported for STD- versus PRC-based reconstructions (stress total perfusion deficits, 0.84 vs. 0.83 [P = 0.49]; ischemic total perfusion deficits, 0.87 vs. 0.87 [P = 0.80]). Conclusion: PRC significantly enhances SNR and CNR compared with STD without affecting the diagnostic accuracy of the scans. Given the significantly improved image quality, PRC may be recommended for MPI using 82Rb PET/CT clinical-routine-assessment interpretation of TPD.
More Related Videos
09:55Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...