A comprehensive analysis of MRI-related Cochlear implant adverse events reported by FDA's manufacturer and user

Amy Lin1, Arjun K Menta2, Syed Ameen Ahmad2

  • 1The Ohio State University College of Medicine Columbus Ohio USA.

Abstract

Insights

MRI-conditional cochlear implants (CI) significantly reduced MRI-related adverse events. Devices with diametric magnets show no reported complications, indicating improved safety for patients undergoing MRI scans.

Area of Science:

  • Biomedical Engineering
  • Medical Device Safety
  • Radiology

Background:

  • Magnetic Resonance Imaging (MRI) poses risks for cochlear implant (CI) recipients due to device magnets.
  • Adverse events, including dislodgement and pain, have been reported with traditional CIs during MRI scans.
  • The development of MRI-conditional CIs aims to mitigate these risks.

Purpose of the Study:

  • To identify and analyze MRI-related adverse events associated with cochlear implants (CI) using FDA MAUDE database.
  • To determine if cochlear implants with diametric magnets are linked to a reduction in MRI-related adverse events.

Main Methods:

  • Retrospective analysis of MRI-related adverse event reports from the FDA MAUDE database (01/01/2013–01/29/2023).
  • Manual extraction of outcomes, interventions, device problems, electrode designs, and patient problems.
  • Statistical analysis using R to group and analyze complications from three major CI manufacturers.

Main Results:

  • A total of 668 MRI-related CI complications were reported (2.1% of 32,446 reports).
  • Common issues included device dislodgment/migration (73.3%) and patient pain (31.6%).
  • A significant decrease in median adverse events per year was observed after the introduction of MRI-conditional CIs (p=0.014).

Conclusions:

  • The introduction of MRI-conditional cochlear implants has demonstrably reduced MRI-associated adverse events.
  • No MRI-related complications have been reported for MRI-conditional CI devices from the three major US manufacturers.