Does protruding headless cannulated screw reduce fixation stability in tension band wiring technique for patella
Mehmet Maden1, Omer Berkay Bayraktar2, Tayfun Bacaksiz2
1Department of Orthopaedics and Traumatology, Izmir Ataturk Training and Research Hospital, Izmir, Turkey. mhmtmdn@gmail.com.
Background:
The selection of an implant is a critical factor in the surgical treatment of patella fractures due to the risk of various complications, such as non-union, implant failure, and irritation. The present study evaluated and compared the biomechanical strength of headless cannulated screws about screw length using the tension band wiring technique.
Methods:
Forty-eight sawbone patellas with transverse fractures were divided into three fixation groups based on the screw length used in tension band wiring. Overall, three different fixation groups were determined: Group 1 (recessed headless cannulated screw fixation), Group 2 (full-length headless cannulated screw fixation), and Group 3 (protruding headless cannulated screw fixation). A setup was used to simulate a knee with a flexion angle of 60 degrees. Specimens underwent biomechanical testing under axial traction (static test) and cyclic loading (dynamic test). Displacements at 300 Newtons (N), loads at 2 millimetres (mm) displacement, and failure loads were documented for each sample in the static test. In the dynamic test protocol, 10,000 repetitive cycles were performed under physiological load between 100 and 300 N, and final displacements were recorded.
Results:
There were significant differences in the loads achieved at 2 mm displacement levels, and Group 3 demonstrated lower force values compared to other constructs in the static test (P = 0.003). All groups revealed similar displacements at 300 N and failure load values under axial traction. In the dynamic test, Group 3 had significantly higher fracture displacement under cyclic loading compared to the other specimens (P = 0.006).
Conclusions:
This study found headless cannulated screws for transverse patella fracture fixation provide sufficient stability; however, protruding headless screws reduce the fixation strength. Recessed or full-length screws may improve stability and bony healing, potentially preventing complications in patella fractures.
Level Of Evidence:
Biomechanical study N/A.
More Related Videos
05:34A Mini-Invasive Internal Fixation Technique for Studying Immobilization-Induced Knee Flexion Contracture in Rats
Published on: May 20, 2019
04:19Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
Published on: November 8, 2024
Related Concept Videos
Cable Subjected to a Distributed Load
Normal Strain under Axial Loading
Plastic Behavior
Relation between Poisson's ratio, Modulus of Elasticity and Modulus of Rigidity
Plastic Deformations
Plastic Deformations