Technique selection and technical developments for 2D dual-energy subtraction angiography on an interventional C-arm
Ethan P Nikolau1, Joseph F Whitehead1, Martin G Wagner2
1Department of Medical Physics, University of Wisconsin - Madison, Madison, Wisconsin, USA.
Background:
Dual-energy (DE) x-ray image acquisition has the potential to provide material-specific angiographic images in the interventional suite. This approach can be implemented with novel detector technologies, such as dual-layer and photon-counting detectors. Alternatively, DE imaging can be implemented on existing systems using fast kV-switching. Currently, there are no commercially available DE options for interventional platforms.
Purpose:
This study reports on the development of a prototype fast kV-switching DE subtraction angiography system. In contrast to alternative approaches to DE imaging in the interventional suite, this prototype uses a clinically available interventional C-arm equipped with special x-ray tube control software. An automatic exposure control algorithm and technical features needed for such a system in the interventional setting are developed and validated in phantom studies.
Methods:
Fast kV-switching was implemented on an interventional C-arm platform using software that enables frame-by-frame specification of x-ray tube techniques (e.g., tube voltage/kV, pulse width/ms, tube current/mA). A real-time image display was developed on a portable workstation to display DE subtraction images in real-time (nominal 15 frame/s). An empirical CNR-driven automatic exposure control (AEC) algorithm was created to guide DE tube technique selection (kV pair, ms pair, mA). The AEC model contained a look-up table which related DE tube technique parameters and air kerma to iodine CNR, which was measured in acrylic phantom models containing an iodine-equivalent reference object. For a given iodine CNR request, the AEC algorithm estimated patient thickness and then selected the DE tube technique expected to deliver the requested CNR at the minimum air kerma. The AEC algorithm was developed for DE imaging performed without and with the application of anti-correlated noise reduction (ACNR). Validation of the AEC model was performed by comparing the AEC-predicted iodine CNR values with directly measured values in a separate phantom study. Both dose efficiency (CNR2/kerma) and maximum achievable iodine CNR (within tube technique constraints) were quantified. Finally, improvements in DE iodine CNR were quantified using a novel variant to the ACNR approach, which used machine-learning image denoising (ACNR-ML).
Results:
The prototype system provided a continuous display of DE subtraction images. For standard DE imaging, the AEC-predicted iodine CNR values agreed with directly measured values to within 3.5% ± 1.6% (mean ± standard deviation). When ACNR was applied, predicted iodine CNR agreed with measurement to within 2.1% ± 3.3%. AEC-generated DE techniques were typically (low/high energy): 63/125 kV, 10/3.2 ms, with varying mA values. When ACNR was applied, dose efficiency was increased by a factor of 9.37 ± 2.08 and maximum CNR was increased by a factor of 3.29 ± 0.21 relative to DE without denoising. Application of ACNR-ML yielded a greater increase in both the dose efficiency (16.11 ± 2.99) and maximum CNR (4.46 ± 0.31) compared to DE without denoising.
Conclusion:
A prototype DE subtraction angiography system using fast kV-switching was implemented on a clinically available interventional C-arm platform without modification of system hardware. The technical features presented in this work include a real-time image display, noise-reduction strategies, and a CNR-driven AEC algorithm. This prototype system demonstrates the feasibility of 2D dual-energy imaging for image-guided interventions.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
X-ray Imaging
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...