Related Experiment Video
Updated: May 29, 2025

A Whole Cell Bioreporter Approach to Assess Transport and Bioavailability of Organic Contaminants in Water Unsaturated Systems
Published on: December 24, 2014
Impact of glass packaging leaching on the uncertainty of inorganic standard solutions
Juliana Serna Saiz1, Diego Alejandro Ahumada Forigua2, Paula Lucía Sánchez García3
1Instituto Nacional de Metrología de Colombia, Ak. 50 26 - 55 Int. 2, 111321, Bogotá, Colombia. jserna@inm.gov.co.
Abstract:
The stability of certified reference materials (CRMs) is essential for ensuring reliable measurements and maintaining metrological traceability. This study presents the first comprehensive evaluation of leaching-induced uncertainty using advanced modeling approaches. We analyzed the leaching behavior of elements such as B, Na, Mg, Ca, and Zn from glass ampoules under controlled conditions, including exposure to 5% nitric acid at 50 °C for 28 days. The analysis detected significant concentrations of these elements, while others like Fe, Cd, and Pb were not detected. By applying logarithmic regression models, we predicted leaching behavior over a 365-day CRM shelf life. The results indicated that, although leaching led to concentration increases-26% for B, 28% for Na, and 15% for Mg-the associated uncertainty remained low, around 0.01% for standard concentrations of 1000 mg kg . These findings highlight the reliability of glass ampoules as a packaging option for CRMs and establish a methodological foundation for incorporating leaching effects into uncertainty estimations. This approach supports the development of improved storage protocols and enhances the long-term stability and accuracy of CRMs, contributing to advancements in the chemical metrology area.
More Related Videos
08:57VacuSIP, an Improved InEx Method for In Situ Measurement of Particulate and Dissolved Compounds Processed by Active Suspension Feeders
Published on: August 3, 2016
08:18A Small-Scale Setup for Algal Toxicity Testing of Nanomaterials and Other Difficult Substances
Published on: October 10, 2020
Related Concept Videos
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Standard Solutions
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Data Validation
Key parameters for method validation include: