Spectral performance for iodine quantification of a dual-source, dual-kV photon counting detector CT
Tim Winfree1, Kevin Treb1, Cynthia McCollough1
1Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.
Background:
Multi-energy CT (MECT) enables quantification of material concentrations by measuring linear attenuation coefficient line integrals with multiple x-ray spectra. Photon counting detector (PCD)-CT utilizes a detector-based approach for MECT that can suffer from substantial spectral overlap, resulting in amplified material quantification noise. Dual-source dual-kV approaches for MECT are currently utilized in some energy-integrating detector (EID)-CT systems and can potentially be utilized with PCD-CT for improved spectral separation.
Purpose:
To evaluate the iodine quantification performance of a novel dual-source (DS)-PCD-CT scan mode and compare to single-source (SS)-PCD-CT and DS-EID-CT.
Materials And Methods:
A 30 cm × 40 cm solid water phantom with four iodine inserts (2, 5, 10, and 15 mg/mL) was scanned with the three spectral modalities: SS-PCD-CT with two energy thresholds, DS-PCD-CT (90/Sn150 kV), and DS-EID-CT (90/Sn150 kV). For each modality, full-dose (12 mGy) and half-dose scans were acquired, and images were reconstructed with filtered back-projection using a quantitative (Qr40) kernel. When scanning in a DS configuration, the total radiation dose budget is split between two detectors, increasing the strength of a signal-dependent filter compared to a SS acquisition. To account for this effect, the modulation transfer function (MTF) for each modality was measured from a 0.05 mm tungsten wire. A linear spatial filter was applied to the SS images to match their MTF to that of the DS images. The resulting high- and low-energy images were input into an image-domain least squares material decomposition algorithm with iodine and water as the two basis materials. Iodine quantification accuracy and noise measured from the iodine basis images were used as figures of merit, and t-tests used to compare between modalities.
Results:
The 10% MTF cutoffs were 0.56, 0.57, 0.60, and 0.57 lp/mm for DS-EID-CT, DS-PCD-CT, SS-PCD-CT, and SS-PCD-CT after MTF-matching, respectively, with the SS-PCD-CT MTF cutoff dropping to 0.58 lp/mm at half-dose. Without accounting for the signal-dependent filter by matching the MTFs, the noise in iodine material basis images from SS-PCD-CT was 10% higher than that of DS-EID-CT. After matching the MTFs, the noise in the SS-PCD-CT iodine image was 9%-22% lower than that of the DS-EID-CT. The lowest iodine image noise was from the DS-PCD-CT, which was 39%-41% lower than the DS-EID-CT. The DS-PCD-CT noise magnitude was significantly different from the other modalities. Mean iodine quantification accuracy across all measured concentrations was within 5% for all modalities at full dose, but was only below 5% for the DS-PCD-CT at half-dose.
Conclusions:
SS-PCD-CT with two energy thresholds outperformed DS-EID-CT in terms of image noise in iodine basis images when spatial resolution was matched. DS-PCD-CT gave the lowest noise due to the combination of PCD technology and improved spectral separation from the different x-ray spectra.
More Related Videos
11:27Studying Soft-matter and Biological Systems over a Wide Length-scale from Nanometer and Micrometer Sizes at the Small-angle Neutron Diffractometer KWS-2
Published on: December 8, 2016
07:48High Spatial Resolution Chemical Imaging of Implant-Associated Infections with X-ray Excited Luminescence Chemical Imaging Through Tissue
Published on: September 30, 2022
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
UV–Vis Spectrometers
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Atomic Emission Spectroscopy: Instrumentation