Usefulness of Blood Flow Measurement Device Using Bioelectrical Impedance Plethysmography in Lower-Extremity Artery
Shigeo Horinaka1, Masashi Sakuma1, Yutaka Yonezawa1
1Department of Cardiovascular Medicine, Dokkyo Medical University Tochigi Japan.
Background:
Bioelectrical impedance plethysmography (IPG) for measuring human body fraction and disease has been progressing in the past half-century, and few studies have reported lower-extremity arterial disease (LEAD) in recent years.
Methods And Results:
The present study enrolled patients who underwent examinations for LEAD. IPG with venous occlusion was performed, and flow volumes were compared with those measured using Doppler duplex ultrasonography, the ankle-brachial index (ABI), and assessments of arterial stenosis and collaterals using computed tomography and/or magnetic resonance angiographies. Fifty patients suspected of LEAD were enrolled; 15 had no arterial stenosis and 35 had LEAD. Arterial blood flow volume (BFV) was assessed. Although the area under the curve for IPG-BFV and Doppler-BFV in the popliteal artery with arterial stenosis were similar, IPG-BFV exhibited better diagnostic accuracy than Doppler-BFV (accuracy 0.765 and 0.694, respectively; McNemar's test P<0.01). In the analysis of covariance with IPG-BFV adjustment, Doppler-BFV was significantly lower in patients with LEAD (ABI<0.9), and morphological arterial stenosis, particularly in those with collaterals than in those without (F-test P<0.05, respectively).
Conclusions:
IPG-BFV could have a better ability to discern the presence of arterial stenosis compared with Doppler-BFV and might not be confounded by the presence of collateral circulation when assessing blood flow in the entire lower extremity, which could be an advantage of IPG-BFV.
More Related Videos
09:18Author Spotlight: Assessing the Reliability of Doppler Ultrasound in Measuring Leg Blood Flow
Published on: December 15, 2023
10:21Evaluation of Hydration Status by Bioelectrical Impedance Vector Analysis in Patients with Ischemic Heart Disease Undergoing Exercise Stress Test
Published on: September 22, 2023
Related Concept Videos
Pulse Assessment Sites
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessing Blood pressure in the Leg
Preparation:
Pre-Procedural Guidelines for Assessing Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to...