Ultrasonographic Optic Nerve Sheath Diameter Technical Pitfalls and Imaging Artifacts.
Mohammad I Hirzallah1,2, Aarti Sarwal3, Aaron M Dentinger4
1Departments of Neurology and Neurosurgery, Baylor College of Medicine, Houston, Texas, USA.
Summary
Ultrasonographic optic nerve sheath diameter (ONSD) measurement is a non-invasive method to assess intracranial pressure (ICP). This review details how ultrasound settings and artifacts impact ONSD accuracy.
Area of Science:
- Neuroscience
- Medical Imaging
- Critical Care Medicine
Background:
- Ultrasonographic optic nerve sheath diameter (ONSD) is a valuable non-invasive surrogate for intracranial pressure (ICP) monitoring.
- Accurate ONSD assessment is crucial for timely clinical decision-making in various neurological conditions.
Purpose of the Study:
- To comprehensively review the impact of ultrasound settings on ONSD measurements.
- To identify and explain common imaging artifacts affecting ONSD assessment.
- To discuss intrinsic optic nerve sheath (ONS) properties influencing ultrasonographic evaluation.
Main Methods:
- Review of existing literature on ultrasound physics and ONSD assessment.
- Analysis of how specific ultrasound parameters (gain, dynamic range, frequency, harmonic imaging, focal zones) influence ONSD measurements.
- Categorization and description of imaging artifacts (e.g., shadowing, enhancement, comet tail, speckle, partial volume, refraction).
Main Results:
- Ultrasound settings like gain and frequency can significantly alter ONSD values.
- Acoustic shadowing, enhancement, and beam property artifacts can lead to inaccurate ONSD readings.
- Changes in ONS echogenicity and ONS kinking due to ICP variations present challenges in ONSD assessment.
Conclusions:
- Understanding the influence of ultrasound settings and artifacts is essential for reliable ONSD measurements.
- Careful technique and artifact recognition are critical for accurate non-invasive ICP assessment using ONSD.
- Further standardization of ONSD measurement protocols may improve its clinical utility.