Diagnostic Accuracy of Three Methods of Body Temperature Measurement in Children: A Latent Class Approach
Kunnumpurath G Swapna1,2, Biju George3, Rajamohanan K Pillai1
1School of Public Health, Kerala University of Health Sciences, Trivandrum, IND.
Insights
The forehead thermometer showed the highest accuracy in detecting fever in children aged one to five years, outperforming axillary and tympanic methods. This study used latent class analysis (LCA) to evaluate diagnostic accuracy without a gold standard.
Area of Science:
- Pediatrics
- Medical Diagnostics
- Biostatistics
Background:
- Fever is a common indicator of acute illness in children, necessitating accurate temperature measurement.
- Current methods for pediatric temperature measurement lack a universally accepted gold standard.
- Latent class analysis (LCA) is an emerging statistical method for evaluating diagnostic accuracy in the absence of a reference standard.
Purpose of the Study:
- To evaluate the diagnostic accuracy of axillary, forehead, and tympanic thermometers in children aged one to five years.
- To utilize Latent Class Analysis (LCA) as a reference standard for fever diagnosis.
- To compare the sensitivity and specificity of different thermometer types in pediatric populations.
Main Methods:
- A cross-sectional study involving 728 children aged one to five years in a tertiary care center.
- Utilized digital axillary and infrared dual-mode (forehead and tympanic) thermometers for temperature measurements.
- Employed Latent Class Analysis (LCA) to establish fever and no-fever classes for diagnostic evaluation.
Main Results:
- Latent class analysis identified two groups: 8.55% with fever and 91.45% without fever.
- The forehead thermometer demonstrated the highest diagnostic accuracy (98.99% sensitivity, 100% specificity).
- Axillary thermometer showed high accuracy (96.97% sensitivity, 97.67% specificity), while the tympanic thermometer had high sensitivity (99.7%) but low specificity (67.44%).
Conclusions:
- The forehead thermometer is the most accurate method for diagnosing fever in children aged one to five years.
- The axillary thermometer is a reliable alternative, offering high diagnostic accuracy.
- The tympanic thermometer, despite high sensitivity, is less reliable due to a significant false-positive rate.
Abstract:
Introduction Fever is a common manifestation of acute illness among children, and it is essential to measure body temperature accurately in pediatric clinical practice. Various methods are in use, but no gold standard exists for body temperature measurement among this population. Latent class analysis (LCA) is increasingly used to assess diagnostic accuracy in the absence of a gold standard. LCA is a method that identifies unobserved groups in populations, allowing diagnostic evaluation even without a reference standard. This study aimed to assess the diagnostic accuracy of the axillary, forehead, and tympanic thermometers in children (one to five years) using LCA. Methods A cross-sectional study was done to determine the diagnostic accuracy of axillary, forehead, and tympanic thermometers in diagnosing fever among children with LCA as a reference. The digital axillary thermometer and the infrared dual-mode (forehead and tympanic) thermometer were used for the measurements. The study was conducted among 728 in a tertiary care center in Kerala, India. Children were recruited from the pediatric medical wards after 24 hours of admission. Results The study sample consisted of 728 children aged one to five years. LCA had identified two latent classes with a probability of 8.55% for fever and 91.45% for no fever. The sensitivity and specificity of the methods were: axillary method 96.97% and 97.67%, forehead thermometer 98.99% and 100%, and tympanic thermometer 99.7% and 67.44%. The probability of misclassification was minimal (0.10%), and a strong association between the test results and the latent class was observed. Conclusions The forehead thermometer demonstrated the highest diagnostic accuracy, followed closely by the axillary thermometer. The tympanic thermometer, though highly sensitive, exhibited a relatively high false-positive rate.
More Related Videos
04:54Using a Combination of Indirect Calorimetry, Infrared Thermography, and Blood Glucose Levels to Measure Brown Adipose Tissue Thermogenesis in Humans
Published on: June 2, 2023
05:12Portable Thermographic Screening for Detection of Acute Wallenberg's Syndrome
Published on: September 19, 2019
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...