Optimizing gas chromatography parameters for thermally unstable dicyclopentadiene mixtures analysis
Azeem Khan1, Mahak Dhiman1, Ankit Mishra2
1Bio-fuels Division, CSIR-Indian Institute of Petroleum, Haridwar Road, Mohkampur, Dehradun 248005, India; Academy of Scientific and Innovative Research (AcSIR), Ghaziabad 201002, India.
Abstract:
Dicyclopentadiene (DCPD) is a versatile compound used in various applications, including resins, polymers, and high-energy-density (HED) fuels, such as exo-tetrahydrodicyclopentadiene (exo-THDCPD). DCPD reaction mixtures are typically analyzed using gas chromatography (GC), but this process can be challenging due to its thermal instability. At temperatures above 150 °C, it can undergo a reverse Diels-Alder reaction (RDAR), decomposing into cyclopentadiene (CPD). This decomposition can lead to significant errors in quantitative measurements, including conversion and yield. To address this issue, we conducted GC analyses at temperatures exceeding the RDAR threshold to investigate DCPD dissociation under various conditions, including GC inlet temperatures, gas flow rates, and solvents. Our study reveals that at inlet temperatures above 200 °C, accurately determining DCPD conversion is extremely difficult. Additionally, we report that the flow rate of the carrier gas has a negligible impact on the DCPD dissociation, while the choice of solvent significantly affects the detection of the CPD formed. Among the three solvents examined, dichloromethane (DCM) was found to be the most effective for detecting dissociated CPD.
More Related Videos
Related Concept Videos
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Gas Chromatography: Types of Detectors-II