Are urine dipsticks reliable indicators of hematuria in blunt trauma patients?
Abstract:
Dipstick results in patients with blunt trauma have not been demonstrated to be predictive of the amount of microscopic hematuria. Our study examined the relationship between dipstick results and the degree of microscopic hematuria in 185 patients admitted to our emergency department following blunt trauma. Urine samples of 169 were dipsticked; the remaining 16 were grossly bloody. A wide range of microscopic hematuria results was obtained for each dipstick value. Of the 80 patients with negative dipsticks, 15 (18.7%) actually had microscopic hematuria. Conversely in 16 of 32 patients (50%) with a "trace" positive dipstick value, no microscopic hematuria was present. Similarly the dipstick value of 1+ had three of 14 (21%) false positives. False positives were recorded seven of 26 times (27%) for 2+, and three of 17 times (18%) for 3+ dipstick readings. We conclude that dipstick results do not correlate well with the presence or absence of hematuria in patients with blunt trauma. Microscopic urinalysis is a better test on which to base the decision to perform further diagnostic tests in the evaluation of the urinary tract.
Insights
Urine dipstick tests are unreliable for detecting microscopic hematuria in blunt trauma patients. Microscopic urinalysis is a more accurate method for guiding further diagnostic decisions.
Area of Science:
- Urology
- Emergency Medicine
- Clinical Pathology
Background:
- Blunt trauma necessitates evaluation for urinary tract injury.
- The predictive value of urine dipstick tests for microscopic hematuria in trauma is uncertain.
Observation:
- A study evaluated 185 blunt trauma patients, analyzing urine dipstick results against microscopic hematuria levels.
- 169 samples underwent dipstick testing; 16 had gross hematuria.
Findings:
- Dipstick results showed poor correlation with microscopic hematuria.
- 18.7% of "negative" dipsticks had microscopic hematuria.
- Significant false positive rates were observed for "trace" (50%), 1+ (21%), 2+ (27%), and 3+ (18%) dipstick values.
Implications:
- Urine dipstick tests are not reliable for ruling in or out hematuria in blunt trauma.
- Microscopic urinalysis is recommended for accurate assessment and guiding further diagnostic workup.