Related Experiment Video
Updated: May 28, 2025

Fabrication of Mechanically Tunable and Bioactive Metal Scaffolds for Biomedical Applications
Published on: December 8, 2015
Reduction of stress-shielding and fatigue-resistant dental implant design through topology optimization and TPMS
Hüray Ilayda Kök1, Miriam Kick1, Osman Akbas2
1Leibniz University Hannover, Institute of Continuum Mechanics, Hannover, Germany.
Abstract:
To improve longevity and performance of dental implants by reducing stress-shielding, a modification of the internal structure of the implant can be a solution. In this paper the inner design of the implant is generated either using a topology optimization approach or an approach based on TPMS lattice structures. These approaches aim to maintain long-term stability and to reduce stress-shielding. For both approaches, the mechanostat model was applied to investigate the influence of the inner structure to the surrounding bone tissue and compare the standard uniform implant. For the investigation an ANSYS model was used with material parameters obtained from a mechanical test of additively manufactured Ti6Al4V. Compared to the uniform implant, the topology-optimized implant showed 20% less stress-shielding, and the implant with triply periodic minimal surface structures (TPMS) showed 15% less stress-shielding. Further, the long-term-stability was investigated by introducing a high-cycle fatigue material model. Despite a change in the internal structure and a 45% reduction in the mass of the topology-optimized implant, the cycle numbers specified in the DIN EN ISO 14801 standard were fulfilled.
Related Concept Videos
Stress Concentrations in Circular Shafts
Stress Concentrations
The stress...
General State of Stress
Specifically, consider a tetrahedral element where one face, labeled XYZ, is perpendicular to the line OA, and the remaining faces align with the coordinate axes with point O as the origin. At any point, such as point O, the stress tensor can be used to determine the stress...