O-cresol Concentration Online Measurement Based On Near Infrared Spectroscopy Via Partial Least Square Regression
Published on: November 8, 2019
Benchtop and different miniaturized near-infrared spectrometers application study: Calibration transfer and 2D-COS
Jianan Yue1, Hui Zhang2, Lele Gao1
1NMPA Key Laboratory for Technology Research and Evaluation of Drug Products, School of Pharmaceutical Sciences, Cheeloo College of Medicine, Shandong University, Jinan 250012 China; Key Laboratory of Chemical Biology (Ministry of Education), Shandong University, Jinan 250012 China.
Abstract:
The demand for miniaturized near-infrared (NIR) spectrometers has surged due to their potential for in-situ analysis. However, their predictive accuracy has not yet matched that of traditional benchtop instruments. This study evaluates the effectiveness of rapid quantitative moisture analysis in hydroxypropyl methylcellulose (HPMC) using a benchtop spectrometer, Antaris II from Thermo Fisher Scientific Inc., and five miniaturized spectrometers (MicroNIR 1700 from Viavi Solutions, OTO-SW2540 from OtOPhotonics, IAS DLP 1700 from Dallas, NIRONE Sensor 2.2 from Helsinki, and NIRS M1800 from Alian Optoelectronics). This study employed an Improved Principal Component Analysis (IPCA) transfer method to standardize spectra from the diverse miniaturized NIR spectrometers, facilitating calibration transfer across different spectroscopic technologies. The benchtop (Antaris II) delivered the most superior results, indicating that miniaturized spectrometers must refine their methodologies to approach the predictive performance of benchtop counterparts. Further, this work conducted a two-dimensional correlation spectroscopy (2D-COS) analysis on the spectra from various spectrometers. This analysis bolstered the partial least squares regression (PLSR) model, highlighting discrepancies between miniaturized and benchtop spectrometers and deepening understanding of the factors influencing the PLSR models. The IPCA leverages the benchtop model to enhance the precision and reliability of miniaturized NIR spectrometers. This innovative and versatile research approach aims to further optimize the performance of miniaturized NIR spectrometers for specific applications, representing a significant step forward in their development.
More Related Videos
10:25Construction of Models for Nondestructive Prediction of Ingredient Contents in Blueberries by Near-infrared Spectroscopy Based on HPLC Measurements
Published on: June 28, 2016
10:22Split Point Analysis and Uncertainty Quantification of Thermal-Optical Organic/Elemental Carbon Measurements
Published on: September 7, 2019
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...