On-Site Evaluation and Improvement Strategies of Radiation Occupational Hazard Prevention and Control Effectiveness
Hong-Xia Xie1, Ming-Qing Zhang1, Zhi-Xiang Li1
1Jinan Hospital, Jinan, Shandong, 250013, People's Republic of China.
Objective:
To evaluate the control effectiveness of medical institution construction projects, and to summarize and analyze the radiation protection management status and improvement strategies of relevant medical institutions.
Methods:
A total of 40 medical institutions in our city were evaluated for control effectiveness through measures such as data research, on-site investigations, equipment quality testing, and radiation health protection inspections.
Results:
The compliance rates of personnel configuration, verification of protective measures, and radiation protection management and emergency response were 95.0%, 67.5%, and 70.0%, respectively. Compared to earlier evaluation periods (eg, before the implementation of new DR performance testing standards, where compliance rates were below 60%), there has been a marked improvement in compliance with performance and protection testing, particularly after the introduction of updated evaluation criteria. The first-pass rates of performance testing for DSA, DR, CT, and dental equipment were 100%, 84.0%, 92.0%, and 100%, respectively. The first-pass rates of radiation protection inspection for related equipment rooms were 100%, 100%, 92.0%, and 100%, respectively. New DR performance testing standards introduced specialized testing items, such as dark noise, detector dose indication (DDI), and signal transmission characteristics (STP), which presented initial challenges due to the unavailability of pre-processing images in some manufacturers' products. Additionally, higher monitoring values were identified at doors, door gaps, and cable penetration points in equipment rooms. Regarding radiation protection management and emergency response, issues such as overly rigid emergency response plans, insufficient personal dose management, and inadequate occupational health examinations remain, requiring systematic adjustments.
Conclusion:
At present, the awareness of radiation hazard prevention and control in medical institutions has been improved. Compared to earlier periods of testing and evaluation, there has been a significant improvement in the degree of compliance with performance and protection testing. Medical institutions have strengthened equipment annual inspections, quality control, and other management work, further enhancing the level of radiation protection management.
Related Concept Videos
Healthcare Associated Infections II: Preventive Measures
The best practices for preventing healthcare-associated infections include hand hygiene, patient risk...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Radiological Investigation I: X-ray and CT
Standard Precaution
Hand hygiene is the most crucial means to prevent the transmission of disease. Employers are legally required to provide their workers with personal protective equipment (PPE) to minimize exposure or contact with...
Nuclear Power
Nuclear Fuels
Nuclear fuel consists of a fissile isotope, such as uranium-235, which must be present in sufficient quantity to provide a...
Biological Effects of Radiation