Single-Cell/Particle Sample Introduction Device for Mass Cytometry Based on Gas-Driven Flow Focusing.
Huaiyi Chen1,2, Jinhui Liu3, Yu Zhao4
1State Key Laboratory of Precision Measurement Technology and Instruments, Department of Precision Instrument, Tsinghua University, Beijing 100084, China.
A new device for mass cytometry analysis prevents clogging using a gas-driven flow focusing protocol. This plug-and-play system ensures smooth sample introduction for long-term single-cell analysis without compromising detection performance.
Area of Science:
- Biotechnology
- Analytical Chemistry
- Single-cell analysis
Background:
- Mass cytometry offers advanced multi-marker analysis but suffers from sample introduction issues like clogging.
- Conventional nebulizers in mass cytometry systems are prone to clogging, hindering long-term sample analysis.
Purpose of the Study:
- To develop a novel, clog-free sample introduction device for mass cytometry.
- To improve the reliability and ease of use of mass cytometry systems for single-cell and single-particle analysis.
Main Methods:
- Developed a plug-and-play device utilizing a gas-driven flow focusing protocol with a virtual hole.
- Achieved single-cell monodispersion via liquid jet breakup, bypassing conventional nebulization.
- Employed high-speed microscopy to study liquid jet dynamics and used inductively coupled plasma mass spectrometry for validation.
Main Results:
- The developed device successfully generated single-cell monodispersion without clogging.
- Comparative analysis showed consistent results between the new device and conventional nebulizers.
- The device demonstrated high universality and compatibility with mass cytometry and ICP-MS instruments.
Conclusions:
- The novel flow focusing device provides a reliable, clog-free solution for mass cytometry sample introduction.
- This innovation enhances the practicality of long-term, high-throughput single-cell analysis.
- The device maintains analytical performance while significantly reducing operational challenges.