Long-term quantitative stability of a first-generation dual-source photon-counting CT
Leening P Liu1, Pouyan Pasyar1, Olivia F Sandvold1
1Department of Bioengineering and Department of Radiology, Perelman School of Medicine, Philadelphia, PA, USA.
Photon-counting CT (PCCT) shows excellent temporal stability over two years, ensuring reliable quantitative imaging for longitudinal studies. This stability enhances diagnostic confidence and expands clinical applications in radiology.
Area of Science:
- Medical Imaging
- Radiology
- Photon-Counting CT
Background:
- Photon-counting CT (PCCT) offers high spatial resolution and quantitative capabilities for advanced imaging.
- Temporal stability of PCCT is crucial for accurate longitudinal studies, distinguishing biological changes from scanner variations.
Purpose of the Study:
- To assess the temporal stability of a first-generation dual-source PCCT over a two-year period.
- To evaluate the impact of software/hardware updates on PCCT quantitative accuracy and noise.
Main Methods:
- A phantom was scanned weekly for two years using single-source and dual-source PCCT modes.
- Virtual monoenergetic images (VMIs) and iodine density maps were analyzed for changes in relative error and noise.
Main Results:
- VMIs showed minimal variation (0.03-0.08%) and stable noise levels (max 4 HU) over time, with improvements in dual-source mode post-updates.
- Iodine density maps remained stable (variation up to 0.1 mg/mL), with enhanced quantification in dual-source mode.
- Relative error in iodine quantification matched between single-source and dual-source modes after updates (-0.04 and -0.02 mg/mL).
Conclusions:
- PCCT demonstrates robust temporal stability, crucial for reliable quantitative metrics in longitudinal studies.
- Improved quantification accuracy, especially in dual-source mode, enhances confidence in diagnostic radiology workflows.
- The findings support the expanded clinical application of PCCT for precise quantitative imaging.