Concentric Needle Jitter Measurements in Voluntarily Activated Deltoid Muscle in Healthy Control Subjects.
João Aris Kouyoumdjian1, Carla Renata Graca1
1Faculdade de Medicina de Sao Jose do Rio Preto, Sao Jose do Rio Preto, State of São Paulo, Brazil.
Muscle & Nerve
|February 18, 2025
Summary
This study establishes normative data for single-fiber electromyography (concentric needle jitter) in the deltoid muscle. New reference values for mean consecutive difference (MCD) in healthy adults are provided for improved neuromuscular disorder diagnosis.
Area of Science:
- Neurology
- Electromyography
- Muscle Physiology
Background:
- The deltoid muscle is sensitive to neuromuscular abnormalities.
- Normative data for single-fiber electromyography (concentric needle jitter) in the deltoid muscle is currently unavailable.
- This gap hinders accurate diagnosis of conditions like myasthenia gravis.
Purpose of the Study:
- To establish normative electrophysiological data for concentric needle jitter in the deltoid muscle.
- To define reference values for mean consecutive difference (MCD) in healthy adults.
- To provide a baseline for diagnosing neuromuscular junction disorders affecting the deltoid muscle.
Main Methods:
- Assessed jitter using mean consecutive difference (MCD) in the voluntarily activated deltoid muscle of 32 healthy adults.
- Calculated the mean MCD of 20 pairs for each subject.
- Determined overall and individual jitter reference values (outliers) based on pooled data.
Main Results:
- Mean MCD values were 21.15 ± 3.42 μs for men and 19.50 ± 2.34 μs for women.
- Individual jitter reference (outlier) values were 27.84 ± 4.68 μs for men and 25.42 ± 3.19 μs for women.
- No prior deltoid muscle jitter values exist for comparison.
Conclusions:
- Proposed normal mean MCD reference values: <28 μs for men and <26 μs for women.
- Proposed normal outlier limits: <38 μs for men and <32 μs for women.
- These values provide essential benchmarks for clinical electrodiagnostic assessments.