Related Experiment Video
Updated: May 27, 2025

Systematic Hearing Performance Evaluation Process for Adolescents with Cochlear Implantation at Early Ages
Published on: March 24, 2023
Functional Benefits of Behind-the-Ear Processors Compared With Off-the-Ear Processors in Adult Cochlear Implant Users
Craig D Salvador1, Zachary Sinacori, Elizabeth L Camposeo
1Department of Otolaryngology-Head and Neck Surgery Medical University of South Carolina, Charleston, South Carolina.
Objective:
Cochlear implant (CI) devices are fitted with two external processor styles-a behind-the-ear (BTE) or an off-the-ear (OTE) option. Although previous research has predominantly focused on speech recognition abilities between processor styles, the current study aims to examine the potential real-world functional differences between processor types.
Study Design:
Retrospective case-control, matched at a 1:2 ratio.
Setting:
Tertiary Otolaryngology Referral Center.
Patients:
Patients with bilateral sensorineural hearing loss.
Intervention:
Cochlear implantation.
Main Outcome Measures:
Cochlear Implant Quality of Life 35 Profile (CIQOL-35 Profile), and CNC word (CNCw) and AzBio sentence (quiet) recognition.
Results:
A total of 36 patients were included (n = 12 OTE and n = 24 BTE users). The overall study population demonstrated improvements in CNCw ( d = 1.9 [1.3, 2.4]), AzBio sentences in quiet ( d = 2.1 [1.5, 2.6]), and medium-to-large effect sizes for domains of the CIQOL-35 ( d range: 0.5-0.9) after cochlear implantation. Between-group analysis demonstrated that BTE users performed better in CNCw ( d = 0.4 [-0.3, 1.1]) and AzBio quiet ( d = 0.5 [-0.2, 1.2]) than their OTE counterparts. However, there were minimal differences identified between processor types based on CIQOL domain and global scores ( d range: 0.04-0.2).
Conclusion:
BTE users may have a better speech recognition ability than their OTE counterparts. However, CIQOL domain and global scores are comparable between the two processor types. A prospective, randomized controlled trial will be needed to address the limitations of a retrospective analysis.
More Related Videos
07:00Electrically Evoked Stapedius Reflex Measurements in Cochlear Implantation and Its Application in the Postoperative Fitting Process
Published on: June 21, 2024
03:58Enhancing Electrode Location Assessment in Cochlear Implantation via Computed Tomography Image Fusion
Published on: January 17, 2025
Related Concept Videos
The Cochlea
Auditory Pathway
When viewed cross-sectionally, the cochlea reveals the scala vestibuli and scala tympani flanking...