Removal of Ticagrelor by Hemoadsorption with the HA380 Cartridge
Taku Furukawa1, Yugeesh R Lankadeva2,3,4, Ian C Baldwin5
1Preclinical Critical Care Unit, Florey Institute of Neuroscience and Mental Health, The University of Melbourne, Melbourne, Victoria, Australia, taku.furukawa@florey.edu.au.
Introduction:
Hemoadsorption has emerged as a potential intervention for the removal of ticagrelor. We aimed to evaluate the efficacy of the HA380 hemoadsorption cartridge for this purpose.
Methods:
Six healthy adult sheep received 270 mg of ticagrelor via an orogastric tube, followed by hemoadsorption using a HA380 cartridge for a duration of 4 h. The sorbent-based removal ratio, clearance, and mass removal rate were assessed at multiple time points.
Results:
The HA380 cartridge achieved an initial sorbent-based removal ratio of 48.9% (SD 11.8) at 10 min, which declined rapidly to 2.66% (SD 18.5) at 120 min and 0.48% (SD 17.0) at 240 min. Clearance followed a similar trend, starting at 46.1 mL/min (SD 11.4) and decreasing to 0.08 mL/min (SD 16.8) at 240 min. The mass removal rate also dropped significantly over time, from 3.74 ng/min (SD 2.54) at 10 min to near zero at 120 and 240 min.
Conclusion:
HA with the HA380 cartridge can achieve an early 50% adsorption level for ticagrelor. If frequently changed, the HA380 cartridge may serve as a potential option for ticagrelor removal, when clinically indicated.