Noninvasive Determination of Vortex Formation Time Using Transesophageal Echocardiography During Cardiac Surgery
Published on: November 28, 2018
Thoracic Aorta Measurements Using Intraoperative Transesophageal Echocardiography: Validation via Cardiac Magnetic
Lisa Q Rong1, Sena Chun1, Pablo Villar-Calle2
1Department of Anesthesiology, Weill Cornell Medicine, New York, NY.
Objectives:
To examine the association and magnitude of agreement between intraoperative transesophageal echocardiography (TEE) and cardiac magnetic resonance imaging (CMR) on aortic diameter measurements, to guide perioperative decision making for ascending aortic aneurysms.
Design:
This prospective study included patients with an ascending aortic aneurysm undergoing surgical ascending aortic repair using a prosthetic graft between February 2018 and December 2019. Intraoperative TEE in 2-dimensional (2D) and 3-dimensional (3D) views were obtained and assessed relative to preoperative cine-CMR and contrast-enhanced 3D magnetic resonance angiography.
Setting:
Weill Cornell Medicine, a single large academic medical center.
Participants:
Thirty patients undergoing ascending aortic repair using a prosthetic graft (polyethylene terephthalate [Dacron]) without severe aortic valve disease (stenosis or regurgitation).
Interventions:
CMR, TEE, and ascending aortic repair using prosthetic grafts.
Measurements And Main Results:
TEE-derived measurements correlated significantly with CMR at all aortic segments: root (r = 0.94-0.97; p < 0.001 for all), ascending (r = 0.95-0.98; p < 0.001 for all), arch (r = 0.88; p < 0.001), and descending (r = 0.91; p < 0.001). Bland-Altman analyses showed small mean differences between 2D and 3D TEE versus CMR (range, 0.05-0.22 cm and 0.04-0.18 cm, respectively), with narrow limits of agreement at all segments, although TEE underestimated CMR-derived measurements.
Conclusions:
This study demonstrates the feasibility of intraoperative TEE to accurately measure aortic diameters in patients with ascending aortic aneurysms. The excellent correlations with CMR and small mean differences with narrow limits of agreement support intraoperative TEE to guide surgical decision making for possible aortic replacement.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above the...
Assessment of the Cardiovascular System IV: Auscultation
Normal Heart Sounds
S1 (First Heart Sound)-
S1 is made by the closure of the mitral and tricuspid valves (atrioventricular valves), marking the beginning of systole.
S2 (Second Heart Sound)-
S2 is made by the closure of the aortic and pulmonic valves (semilunar valves), marking the end of the systole.