3D-printed template design to improve 125I seed plaque assembly accuracy for uveal brachytherapy
Kaitlyn Calabresi1, Jacqueline Emrich2, Sara Belko3
1Department of Radiation Oncology, Thomas Jefferson University Hospital, Philadelphia, PA 19107.
Purpose:
To demonstrate the utility of a QA program for seed localization, and to design a 3D-printed template to improve the accuracy of seed placement on custom-built 125I eye plaques for uveal brachytherapy.
Methods And Materials:
A seed localization analysis tool (SLAT) was developed in MATLAB to detect variations in seed placement relative to a treatment plan. A flexible seed placement template (3D-FSPT) was designed in CAD and printed using a Formlabs Form-3 3D printer. The 3D-FSPT and SLAT were tested using 3D-printed model eye plaques with nonradioactive seeds arranged following clinically-relevant treatment plans. Five clinical plaques were also evaluated.
Results:
SLAT detected submillimeter scale variations in seed position with 2.3% error relative to the plan's seed coordinates, and with an uncertainty of ± 0.01 mm. The average seed displacement on the model plaques with free-handed seed placement was 1.31 mm (SD = 0.61), and the average seed orientation difference was 5.27 degrees (SD = 4.77). The average seed displacement on the clinical plaques was 0.77 mm (SD = 0.42), and the average seed orientation difference was 4.41 degrees (SD = 3.00). For the clinical plaques, changes in dosimetry to the tumor apex and critical eye structures were within acceptable tolerances. Seed displacement (mm) (p < 0.001) and seed orientation differences (degrees) (p = 0.008) were significantly lower using the template to guide seed placement on the model plaques compared to free-handed seed placement.
Conclusions:
The feasibility of a 3D-FSPT and SLAT is demonstrated for improving seed placement accuracy relative to a treatment plan.