Contrast-Enhanced Subharmonic Aided Pressure Estimation SHAPE Using Ultrasound Imaging with a Focus on Identifying Portal Hypertension
Published on: December 5, 2020
An experimental study on subharmonic-aided pressure estimation with commercial scanners
Yi-Fan Dong1, Li Zhang1, Yao Chen2
1Department of Ultrasound, State Key Laboratory of Complex Severe and Rare Diseases, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
Abstract:
This study evaluates the performance of subharmonic-aided pressure estimation (SHAPE) with ultrasound contrast agents using data obtained from calibration procedures. A commercial scanner of ultrasonography (GE Healthcare Logiq E20, Wuxi, China) was used to perform SHAPE on a flowing contrast agent (Sonazoid) in a phantom setup under controlled microbubble stability and flow velocity conditions. Subharmonic time-intensity curves were collected during the SHAPE calibration procedure for analysis. Subharmonic amplitude of diluted contrast agents exhibited a time-dependent decline but was not affected by velocity. SHAPE sensitivity was measured through reciprocal pressurizing and depressurizing sequences to mitigate the effect of subharmonic decline over time. A wide range of mechanical index (MI) levels within the steady growth phase of the calibration curve showed higher SHAPE sensitivity compared to the traditionally recommended "optimal" MI at the maximum slope. The approximate maximum SHAPE sensitivity was -0.04 dB/mm Hg. Subharmonic amplitude was linearly correlated with pressure at a range of MI levels (R2 > 0.9, p < 0.05) but showed significant variations (approximately 2 dB standard deviation) in the time series. The lowered sensitivity compared to previous reports, combined with the substantial variation in subharmonic amplitude, raises concerns about the accuracy and consistency of SHAPE in clinical applications.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: