In Vitro Application of a Wireless Sensor in Flexion-Extension Gap Balance of Unicompartmental Knee Arthroplasty
Published on: May 5, 2023
Distinctive Polyethylene Damage in Rotating-Platform Total Knee Arthroplasty: A Comparative Retrieval Study in
Tracy M Borsinger1, Nicole D Quinlan2, Douglas W Van Citters3
1Department of Orthopaedics, Dartmouth-Hitchcock Medical Center, Geisel School of Medicine at Dartmouth, Lebanon, New Hampshire; Department of Orthopaedic Surgery, University of Virginia Medical Center, Charlottesville, Virginia.
Mobile-bearing rotating-platform total knee arthroplasty (TKA-RP) instability is linked to a specific postero-lateral polyethylene insert damage pattern. This finding aids in diagnosing TKA-RP failures and improving implant design.
Area of Science:
- Orthopedic Surgery
- Biomedical Engineering
- Materials Science
Background:
- Mobile-bearing rotating-platform total knee arthroplasty (TKA-RP) may be sensitive to asymmetric flexion gaps.
- Improper balancing can lead to polyethylene insert instability in TKA-RP.
- A distinct polyethylene damage pattern associated with TKA-RP clinical failure was investigated.
Purpose of the Study:
- To identify a characteristic polyethylene insert damage pattern in TKA-RP associated with clinical instability.
- To correlate observed damage patterns with biomechanical predictions of dynamic instability.
Main Methods:
- Analysis of nine TKA-RP revisions for symptomatic flexion instability (clinical cohort).
- Random selection of 30 TKA-RPs from a revision retrieval database (registry cohort).
- Inspection and categorization of polyethylene inserts using the modified Hood method.
Main Results:
- Seven of nine clinical cohort inserts showed postero-lateral polyethylene damage, from minor wear to gross deformation.
- Eight of 30 registry cohort inserts exhibited similar postero-lateral damage patterns.
- Instability symptoms were reported by nine registry patients; similar damage was seen in eight.
Conclusions:
- A distinctive postero-lateral polyethylene damage pattern was observed in 15 of 18 TKA-RP inserts with clinical instability.
- This damage suggests posterior condyle subluxation over the insert's peripheral wall.
- Aseptic loosening did not prevent the occurrence of this hallmark damage pattern.