Measurement accuracy and cutoffs for predicting primary aldosteronism diagnosis using Lumipulse® for renin and
Maude Laney1, Appoline Nabi-Bourgois1, Julien Mallart-Riancho2,3,4,5
1Physiology Department, AP-HP, Georges Pompidou European Hospital, Paris, France.
Objectives:
Primary aldosteronism (PA) diagnosis is a multistep process that begins with screening based on the plasma aldosterone-to-renin ratio (ARR). As cut-off values depend on the methods used, the aim of this study was 1) to validate the performances of new aldosterone and renin chemiluminescent immuno-assays using the Lumipulse® analyzer (Fujirebio®) and then 2) to define the criteria to distinguish PA patients from those with essential hypertension (EH).
Methods:
This monocentric retrospective study included 297 patients admitted to Georges Pompidou European Hospital, France, between January 2021 and December 2022 for the assessment of hypertension etiology. We evaluated analytical performances and compared Lumipulse® results for plasma renin (PRC) and plasma aldosterone (PAC) concentrations (n = 196) as well as 24-hour urine aldosterone excretion (24UA) (n = 201) to the reference methods (PRC: Liaison XL® Diasorin®; PAC and 24UA: LC-MS/MS). Receiver operating characteristic curve analysis was used to define the cut-off values for ARR, PAC, and 24UA to distinguish between PA and EH patients.
Results:
Analytical characteristics were in accordance with requested performances. Our results showed excellent correlations between Lumipulse® and our reference methods. We propose cut-off values of 35 pmol/mIU as the ARR, 260 pmol/L for PAC, and 49 nmol/24h for 24UA for predicting PA diagnosis.
Conclusions:
Lumipulse® (Fujirebio®) is a reliable alternative for measuring PRC, PAC, and 24UA for predicting PA diagnosis with the use of corresponding cut-off values for each criterion.
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pre-Procedural Guidelines for Assessing Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Errors occurring during blood pressure monitoring
Several factors...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...