DNA Flow Cytometry Analysis in Body Cavity Fluids Using Liquid-Based Cytology
Ayumi Ryu1,2,3, Keiichiro Honma2, Sayoko Tsuzaki1,2
1Department of Clinical Laboratory, Osaka International Cancer Institute, Osaka, Japan.
CelVerse (CeV) preservation solution offers more consistent cell proliferation index (CPIx) results for body cavity fluid cytology using LC-1000 compared to Cellprep (CP). CeV is less affected by SubG1% variations, enhancing diagnostic reliability.
Area of Science:
- Cytopathology
- Flow Cytometry
- Biomedical Analysis
Background:
- DNA flow cytometry with LC-1000 analyzes cellular DNA content and proliferation index (CPIx).
- Body cavity fluid cytology requires reliable cell preservation solutions for accurate analysis.
- Evaluating new preservation solutions like CelVerse (CeV) against established ones like Cellprep (CP) is crucial.
Purpose of the Study:
- To evaluate the performance of body cavity fluid cytology using the LC-1000 system.
- To compare the efficacy of CelVerse (CeV) as a cell preservation solution against Cellprep (CP).
- To assess the impact of different preservation solutions on the cell proliferation index (CPIx) and diagnostic accuracy.
Main Methods:
- Compared 43 body cavity fluid samples in CeV and 59 in CP using LC-1000.
- Analyzed cell proliferation index (CPIx) in malignant and benign cases for both CeV and CP groups.
- Assessed the influence of SubG1% on diagnostic performance, measured by area under the curve (AUC).
Main Results:
- Both CeV and CP showed significant differences in CPIx between malignant and benign cases (p < 0.001).
- CelVerse (CeV) demonstrated a higher AUC (0.811) compared to Cellprep (CP) (0.776).
- The difference in SubG1% significantly impacted CP's AUC, while CeV results remained consistent.
Conclusions:
- Cell proliferation index (CPIx) using LC-1000 in CeV or CP samples is a viable objective indicator for body cavity fluid cytology.
- CelVerse (CeV) offers more consistent results in body cavity fluid cytology, unaffected by SubG1% variations.
- CeV shows potential as a superior preservation solution for LC-1000 based body cavity fluid cytology.