Energy Saving in Permanent Cardiac Pacing: Pulse Waveform and Charge Balancing Deserve Consideration
Franco Di Gregorio1, Lina Marcantoni2, Aldo Mozzi3
1Clinical Research Unit, Medico Spa, 35030 Rubano, PD, Italy.
Abstract:
The pacing pulse produced by implantable stimulators can be described as a truncated exponential decay from the starting peak amplitude, corresponding to the discharge of the output stage capacitance (reservoir and isolation capacitors, in series) along the application time. Pulse decay and charge balancing have relevant implications on the ideal setting of a pacing device, as demonstrated by mathematical predictions based on well-acknowledged theoretical statements. Successful stimulation is achieved with minimum energy expense at a pulse duration shorter than the chronaxie time, which represents the upper border of the advisable duration interval. With any start amplitude, the stimulation safety margin can be improved by a duration increase beyond the chronaxie only up to an absolute limit (longest useful duration), which depends on the chronaxie and the pulse time-constant. At the longest useful duration, the threshold start amplitude is at the minimum and cannot decrease any further, though it and the corresponding pulse mean amplitude largely exceed the rheobase. The overall pacing performance is affected, in addition, by the load resistance and the electrode capacitance. Pulse amplitude decay limits the effectiveness of extended duration in implantable stimulators, making short pulses preferable whenever possible. Proper pulse settings based on actual waveform properties can prevent energy waste and reduce pacing consumption, thus prolonging the service life of the stimulator.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Electrophysiology of Normal Cardiac Rhythm
Special considerations while measuring pulse
Cardiac Action Potential
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
The Cardiac Cycle
The Process
Electrical signals—sent from the sinoatrial (SA) node in the right atrial wall to the atrioventricular (AV) node between the right atrium and right ventricle—cause both atria to simultaneously contract. When the signal reaches the AV node, it pauses for approximately a tenth of a second, allowing the atria to contract and...