Feasibility of Coplanar VMAT for Brain Metastases Using Halcyon With Knowledge-based Planning from Non-coplanar Plan.
Yusuke Sakai1, Kazuki Kubo1, Kenji Matsumoto2
1Department of Medical Physics, Graduate School of Medical Sciences, Kindai University, Osaka, Japan.
In Vivo (Athens, Greece)
|February 26, 2025
Summary
This study shows that knowledge-based planning on the Halcyon system is feasible for single-isocenter volumetric modulated arc therapy (SI-VMAT) for brain metastases. The Halcyon system offers clinically acceptable dosimetric performance comparable to the TrueBeam system.
Area of Science:
- Radiation Oncology
- Medical Physics
- Neurosurgery
Background:
- Multiple brain metastases present a challenge for radiation therapy.
- Single-isocenter volumetric modulated arc therapy (SI-VMAT) offers potential benefits for treating multiple brain lesions.
- Knowledge-based planning (KBP) systems can optimize treatment plans.
Purpose of the Study:
- To evaluate the feasibility of coplanar SI-VMAT using a Halcyon system with a KBP model.
- To assess the accuracy of a KBP model for Halcyon-based SI-VMAT.
- To compare the dosimetric performance of Halcyon and TrueBeam systems for SI-VMAT.
Main Methods:
- A RapidPlan (RP) KBP model was developed using TrueBeam SI-VMAT plans.
- The RP model was validated for the Halcyon system using root-mean-square error (RSME) and dose-volume histogram (DVH) analysis.
- Dosimetric outcomes were compared between TrueBeam and Halcyon SI-VMAT plans for 10 patients.
Main Results:
- The RP model demonstrated good accuracy for Halcyon plan prediction, with minor deviations in the low-dose region.
- While statistically significant differences were observed in certain dosimetric parameters (gradient index, conformity index, normal brain dose), they were clinically negligible (<1% disparity).
- The Halcyon system achieved clinically acceptable dosimetric performance comparable to the TrueBeam system.
Conclusions:
- The developed RP model can accurately predict doses for Halcyon SI-VMAT, irrespective of beam configurations.
- Coplanar SI-VMAT with the Halcyon system is a feasible and effective treatment option for multiple brain metastases.
- The Halcyon system provides clinically comparable dosimetric results to the TrueBeam system for SI-VMAT.