Automatic phantom-less calibration of routine CT scans for the evaluation of osteoporosis and hip fracture risk
Wen Li1, Yuanzhi Weng2, Renfei Zong3
1Department of Radiology, The First Affiliated Hospital of Chongqing Medical University, Chongqing, China.
Background/Purpose:
The diagnosis of osteoporosis remains a paramount concern for orthopedic surgeons worldwide. We aim to (1) evaluate the efficacy of automatic phantom-less quantitative computed tomography (PL-QCT) in diagnosing osteoporosis and (2) investigate its clinical value in predicting hip fracture risk.
Methods:
A cohort of 705 patients was included in the study. Hip CT scans from 310 patients and spinal CT scans from 315 patients were analyzed using automatic PL-QCT. The consistency of bone mineral density (BMD) measurement obtained by dual-energy X-ray absorptiometry (DXA), phantom-based QCT (PB-QCT), and automatic PL-QCT was examined through linear regression analysis and Bland-Altman plots. The ability of automatic PL-QCT to predict osteoporosis and hip fracture risk was assessed using ROC analysis.
Results:
Linear regression and Bland-Altman plots demonstrated a high level of agreement between BMD measurements from PL-QCT and those from hip DXA and lumbar PB-QCT. The AUC values for PL-QCT and PB-QCT in diagnosing osteoporosis were 0.903 (95 % CI 0.852-0.955) and 0.900 (95 % CI 0.847-0.953). The AUC values for predicting hip fracture risk, based on femoral neck BMD measured by PL-QCT and DXA, were 0.869 (95 % CI 0.823-0.915) and 0.831(95 % CI 0.778-0.885), respectively. When the femoral neck BMD was combined with the percentage of inter-muscular adipose tissue area, the AUC increased to 0.929 (95 % CI 0.897-0.961).
Conclusion:
Automatic PL-QCT has shown superior performance in predicting hip fracture risk compared to DXA. Furthermore, the novel PL-QCT demonstrates comparable predictive efficacy to that of PB-QCT, suggesting its potential as a valuable tool in clinical practice.
More Related Videos
07:12Semiautomated Longitudinal Microcomputed Tomography-based Quantitative Structural Analysis of a Nude Rat Osteoporosis-related Vertebral Fracture Model
Published on: September 28, 2017
09:02Cortical Bone Assessment Using Ultrasonic Guided Waves: A Reproducibility Study in a Healthy Population
Published on: January 31, 2025
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT