An underwater gas chromatograph for in-situ analysis of organics in deep-sea
Shihao Lu1, Kun Ding2, Jian Zhang3
1Department of Instrumentation & Analytical Chemistry, CAS Key Laboratory of Separation Sciences for Analytical Chemistry, Key Laboratory of Deep-sea Composition Detection Technology of Liaoning Province, Dalian Institute of Chemical Physics, CAS, 457 Zhongshan Road, Dalian, 116023, China; University of Chinese Academy of Sciences, Beijing, 100049, China.
Background:
Deep-sea exploration has emerged as a pivotal area of focus in contemporary geoscience and analysis techniques. For deep-sea analysis, conventional ex-situ water analytical methodologies mean significant economic costs, while also greatly increasing the risk of sample distortion. The solution to this issue lies in in-situ analysis, eliminating the bias from sampling and transportation, and capturing spatial and temporal data simultaneously. Underwater gas chromatograph (GC) has great potential in in-situ analysis of VOCs in deep-sea, but few studies have verified its feasibility.
Results:
In this work, an underwater membrane inlet GC was developed and evaluated. The membrane permeated gas samples were subjected to dehydration, enrichment, thermal desorption, column injection and separation, and thermal conductivity detector detection. The overall dimension of the underwater GC is Φ 210 mm × 770 mm, 21.0 kg in water, and the design maximum working depth is 4500 m. The average power consumption is ≤ 35 W, and the gas consumption is 3 mL/min. The performance of the underwater GC was evaluated in the laboratory tank. The detection limits of ethane, propane, n-butane, benzene, toluene, and p-xylene in water at a 60-min sampling period were 2.4, 1.1, 0.85, 2.9, 1.2, and 0.73 μg/L. The linear range is around 2 orders of magnitude. The underwater GC has been tested twice in the South China Sea at a depth of more than 1000 m to verify its deep-sea in-situ analysis performance.
Significance:
Our work confirms the feasibility of in-situ analysis of VOCs in water using underwater GC not only in the lab but in deep-sea trials. This study also provides a prototype and reference for the design of underwater analytical instruments.
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...