[Method of linear calibration function control according to GOST R ISO 11095-2007 for measuring the mass
M S Zatyamina1,2,3, L A Smirnova1,2, O V Ostrovskij1
1Volgograd State Medical University, Volgograd, Russia.
Sudebno-Meditsinskaia Ekspertiza
|February 28, 2025
Summary
Calibration functions were validated using a control method with standard and internal reference materials (RMs). The system remained in a controlled state, confirming the calibration function
Area of Science:
- Analytical Chemistry
- Metrology
Context:
- Calibration functions require regular validation to ensure accuracy.
- GOST R ISO 11095-2007 provides a framework for method validation.
- Reference Materials (RMs) are crucial for quality control in analytical measurements.
Purpose:
- To implement a control method for validating existing calibration functions.
- To assess the stability and reliability of calibration functions using both standard and internal RMs.
- To ensure compliance with GOST R ISO 11095-2007 standards.
Summary:
- A control method was applied to validate calibration functions for aqueous ethanol solutions.
- Standard RMs and internally developed RMs were used for validation.
- Control charts demonstrated that the system was in a controlled state, requiring no updates to the calibration function.
- Uncertainty assessment was performed during the validation process.
Impact:
- Confirms the long-term stability of calibration functions.
- Ensures the reliability of analytical measurements for ethanol concentration.
- Provides a robust framework for ongoing quality assurance in chemical analysis.
Keywords:
GOST R ISO 11095-2007comparison sampleconcentrationcontrol chartcontrol methodethanoluncertaintyMore Related Videos
Related Concept Videos
Instrument Calibration
147
Instrument calibration is essential for ensuring that instruments produce accurate and consistent results. It is vital in manufacturing, healthcare, testing laboratories, and scientific research. Calibration processes are specific to each instrument and help enhance data accuracy. Each instrument has a unique calibration process tailored to its design and function to improve data accuracy.
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
147
Calibration Curves: Linear Least Squares
1.2K
A calibration curve is a plot of the instrument's response against a series of known concentrations of a substance. This curve is used to set the instrument response levels, using the substance and its concentrations as standards. Alternatively, or additionally, an equation is fitted to the calibration curve plot and subsequently used to calculate the unknown concentrations of other samples reliably.
For data that follow a straight line, the standard method for fitting is the linear...
For data that follow a straight line, the standard method for fitting is the linear...
1.2K
Glassware Calibration
164
Accurate calibration of glassware, such as volumetric flasks, pipettes, and burettes, is essential to ensure accurate measurements in the analytical laboratory. Calibration helps maintain consistency across measurements and prevents errors arising from inaccurate volumes.
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
164
Flame Photometry: Lab
206
In a flame photometer, when a solution like potassium chloride is aspirated into the flame, the solvent evaporates, leaving behind dehydrated salt. This salt dissociates into free gaseous atoms in their ground state. Some of these atoms absorb energy from the flame, leading to their excitation. The excited atoms return to the ground state, emitting photons at characteristic wavelengths. Because only electronic transitions are involved, the resulting emission lines are very narrow. The intensity...
206
Constant Volume Calorimetry
26.8K
Calorimeters are useful to determine the heat released or absorbed by a chemical reaction. Coffee cup calorimeters are designed to operate at constant (atmospheric) pressure and are convenient to measure heat flow (or enthalpy change) accompanying processes that occur in solution at constant pressure. A different type of calorimeter that operates at constant volume, colloquially known as a bomb calorimeter, is used to measure the energy produced by reactions that yield large amounts of heat and...
26.8K
UV–Vis Spectroscopy: Beer–Lambert Law
1.9K
The Beer-Lambert law describes the relationship between absorbance and concentration, which combines the principles established by scientists Johann Heinrich Lambert and August Beer. Lambert's law states that when light passes through a medium, the loss in intensity is directly proportional to the original intensity and the path length of the light. Beer's law proposed that the transmittance of a solution remains constant if the product of concentration and path length is constant. The...
1.9K