Related Experiment Video
Updated: May 24, 2025

Iris Fixation via External Pentagram Suturing
Published on: May 5, 2022
Low-Cost New Glaucoma Drainage Device in Indonesia: One Year Follow-Up of 252 Subjects
Virna Dwi Oktariana Asrory1, William H Morgan2
1Faculty of Medicine Universitas Indonesia, Cipto Mangunkusumo Hospital, Jakarta, Indonesia.
Background:
To evaluate the early efficacy and safety of a new polymethyl methacrylate 200 mm2 non-valved glaucoma drainage device (GDD) designed for Indonesian use over a 12 month period.
Methods:
This was a prospective study following subjects with various forms of refractory glaucoma aged ≥ 7 years old followed for 12 months. The subjects had IOP measurements and biomicroscopy examination preoperatively and at 1, 3, 6 and 12 months timepoints. We noted anti-glaucoma medication use and additional interventions during follow-up. Success was defined as an IOP > 5 and ≤ 21 mmHg on no medications (complete) or medications (qualified).
Results:
There were 252 subjects, 31% (79/252) primary glaucoma and 69% (173/252) secondary glaucoma. All primary glaucoma had previous surgery. The majority of secondary glaucoma was post-vitrectomy and silicone oil insertion 35.6% (62/174). The mean pre-op IOP was 36.8 ± 12.2 mmHg and the 12 months post op IOP was 14.2 ± 6.6 mmHg. The qualified success rate was 89% at 1 year and the complete success was 57%. Mean medication use dropped from 3.5 to 1.3. Certain subtypes (Silicone oil and POAG) appeared to experience greater IOP reduction, however they also had greater preoperative IOP.
Conclusion:
This new GDD has comparable IOP lowering outcomes compared with other non-valved glaucoma drainage devices during 1 year follow up with similar complication rates.
Related Concept Videos
Angle Closure Glaucoma: Treatment
Glaucoma: Overview
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...