Optimization of imaging conditions for technetium-99m dimercaptosuccinic acid single photon emission computed
Haruto Kimoto1,2, Yoshiki Owaki2,3, Kazuya Minamishima3
1Department of Radiology, Tokyo Medical University Hospital, .
Objective:
This study validated the optimal acquisition and reconstruction conditions for technetium-99m dimercaptosuccinic acid single-photon emission computed tomography examinations in neonates.
Methods:
We developed renal phantoms representing neonatal and adult kidneys, derived from clinical data to achieve this. The radioactivity concentration to be encapsulated in the renal part was adjusted according to the Japanese consensus guidelines for the appropriate implementation of pediatric nuclear medicine examinations. The evaluation of optimal acquisition conditions considered factors such as acquisition time and magnification ratio, while the determination of optimal reconstruction conditions focused on the cutoff frequency of the Butterworth filter and the appropriate correction methods. The physical evaluation assessed defect contrast and background coefficient of variation, while the visual evaluation assessed the ability of the images to depict the defect.
Results:
The defect contrast was almost constant after acquisition times of 10 min for neonates and 12 min for adults. For the magnification ratio, the effect was greater for neonates than for adults. In addition, in contrast to adults, the use of resolution recovery significantly decreased scores in visual evaluation in neonates.
Conclusion:
Our study suggests better images can be obtained in neonates with shorter acquisition times than adults. In addition, the optimal reconstruction conditions are shown to differ between neonates and adults.
More Related Videos
15:18Making MR Imaging Child's Play - Pediatric Neuroimaging Protocol, Guidelines and Procedure
Published on: July 30, 2009
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...