Related Experiment Video
Updated: May 24, 2025

Treatment of Liver Metastases Using an Internal Target Volume Method for Stereotactic Body Radiotherapy
Published on: May 8, 2018
Margins to compensate for respiratory-induced mismatches between lung tumor and fiducial marker positions using
Seiya Matsumoto1, Nobutaka Mukumoto2, Tomohiro Ono3
1Department of Advanced Medical Physics, Graduate School of Medicine, Kyoto University, 53 Kawahara-cho, Shogoin, Sakyo-ku, Kyoto 606-8507 Japan.
Background And Purpose:
Tumors and fiducial markers do not always exhibit synchronous motion across different respiratory phases, in a phenomenon called the target localization error (TLE). We determined the margin to compensate for the TLE using four-dimensional computed tomography (4D-CT).
Materials And Methods:
We analyzed data from 21 lung tumor patients with fiducial markers; 11 for TLE determination and 10 for validation. Shifted CT images were generated by aligning the centroids of the fiducial markers in the reference phase of 4D-CT with those in each respiratory phase, and the union of gross tumor volumes (GTVs) was determined ( ). Conversely, variations in GTV centroids across the respiratory phases were calculated, and the 95th percentile of the root mean square error was defined as the TLE. Using this TLE, a GTV with an added TLE ( ) was generated in the reference phase. Subsequently, a treatment plan assuming dynamic tumor tracking (DTT) was created for the planning target volume, derived by adding an isotropic 5 mm margin to , and the dose coverage for was evaluated.
Results:
The TLEs (standard deviations of the root mean square error) were 2.0 (0.8) mm, 2.1(0.7) mm, and 3.2 (1.1) mm in the left - right, anterior - posterior, and superior - inferior directions, respectively. A dosimetric evaluation revealed that did not receive 100 % of the prescribed dose in four of 10 cases owing to artifacts.
Conclusion:
The TLE can be compensated by adding an anisotropic margin to the GTV in the reference phase, a critical consideration in DTT.