Fluorescent Paper Strips for the Detection of Diesel Adulteration with Smartphone Read-out
Published on: November 9, 2018
A Simple Circuit for Time-Resolved Luminescence (TRL) Measurement Instruments: Demonstration Through a
Alim Gulbag1, Michael Huang1, Brian Rong1
1Department of Electrical and Biomedical Engineering, University of Nevada Reno, NV 98557 USA.
None:
Time-resolved luminescence (TRL) measurement is a sensitive detection technique by eliminating sample autofluorescence, but TRL measurement instruments composed of multiple key components (e.g., a rapidly pulsed light excitation source, a time-gated optical detector, and a synchronization module aligning the timing between the light source and the detector) have been sophisticated, expensive, or bulky, which limits their point-of-care or in-field applications. To reduce the cost and complexity of these instruments, in this work, we developed a simple circuit for rapid LED pulsing and accurate timing synchronization, and implemented it in a compact TRL imager with a UV LED as light source and a chopper-coupled smartphone camera as time-gated optical detector. The TRL measurement of this imager using this circuit was successfully validated through an anticounterfeiting application. We believe that this simple circuit can be adopted in the development of low-cost and compact TRL measurement instruments for broad point-of-care or in-field applications.