Objective:
We developed a linear matrix inequality-enabled absolutely stabilizing proportional-integral-derivative control design approach for pharmacological systems applicable to intravenous drug administration.
Methods:
We developed a proportional-integral- derivative control design approach that does not require detailed knowledge of the dose-response relationship other than its sector bound. It repetitively solves a set of linear matrix inequalities, which encapsulate the Lyapunov stability conditions against unknown dose-response relationship, over a broad proportional-integral-derivative gain space. The linear matrix inequality-feasible proportional-integral-derivative gains guarantee the absolute stability of the closed-loop control system against unknown yet sector-bounded dose-response relationship. The proof-of-concept of the approach was shown in silico using intravenous propofol anesthesia as a practical case scenario.
Results:
The in silico evaluation results demonstrated the robustness and performance of the proportional-integral- derivative controllers designed with the proposed control design approach against unknown sector-bounded nonlinear dose-response relationship and parametric uncertainty in the plant dynamics.
Conclusion:
Pending follow-up development and extensive evaluation in various complex intravenous drug administration problems, the proposed approach may find applications in various closed-loop automated intravenous drug administration problems with complex and highly nonlinear dose-response relationships.
Significance:
The proposed control design approach provides a systematic way to absolutely stabilize pharmacological systems against unknown, nonlinear, and time- varying dose-response relationship, perhaps for the first time.
Related Concept Videos
PID Controller
Drug Delivery: Parenteral Route
There are three primary parenteral routes: intravenous (IV), intramuscular (IM), and subcutaneous (SC). The IV route introduces the drug directly into the bloodstream, ensuring immediate action. The IM route...
One-Compartment Open Model for IV Bolus Administration: General Considerations
The drug's presence in the body is defined by an equation representing the difference between the rates of drug entry and exit. Key parameters—elimination rate constant,...
Rational Dosage Regimen: Maintenance Dose and Loading Dose
In most cases, drugs are administered repetitively or infused continuously to maintain a steady-state concentration in the body. At a steady...
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
Two-Compartment Open Model: IV Infusion
The model illustrates the decrease in plasma drug concentration from the central compartment with a specific equation. It shows that under steady-state conditions, the drug's input rate...