Inter-Channel Correlation-Based EMI Noise Removal (ICER) for Shielding-Free Low-Field MRI
Objective:
For low-field magnetic resonance imaging (MRI) in unshielded environment, existing methods have been proposed to eliminate electromagnetic interference (EMI) noise in each single radio-frequency (RF) receive coil. In the present study, we propose to use the EMI information from multiple MRI receive coils collectively in EMI denoising.
Methods:
The proposed method leverages the information of inter-channel correlation, including EMI detectors and RF receive coils to remove EMI noise. Calibration signals from both EMI detectors and RF receivers are concatenated to determine a de-correlation matrix, which is then used to denoise MRI signals.
Results:
Saline phantom and in vivo experiments demonstrated the efficacy of the proposed method in EMI elimination, showing that the proposed method outperformed advanced EMI elimination methods with up to 16.34% improvement in EMI noise removal percentage (NRP) and 1.58dB improvement in signal-to-noise ratio (SNR), along with reduced computational time.
Conclusion:
The proposed method effectively removes EMI noise and shows improved performance by using information from all receive coils.
Significance:
This method allows for the design of multi-receive coils for low-field MRI such as phased-arrays, which have the potential to enhance the performance in noise removal and improve the SNR in MRI signal acquisition.