Optical 14C Tracing for Biological and Pharmaceutical Applications Using Two-Color Cavity Ringdown Spectroscopy
Jun Jiang1, David Baliu-Rodriguez2, A Daniel McCartt1
1Center for Accelerator Mass Spectrometry, Lawrence Livermore National Laboratory, Livermore, California 94550, United States.
Laser-based Carbon-14 (14C) quantitation using two-color cavity ringdown spectroscopy (2C-CRDS) offers a sensitive, affordable alternative to AMS. This mid-IR laser technique achieves sub-fCi quantitation for 14C tracing in biological and pharmaceutical studies.
Area of Science:
- Analytical Chemistry
- Spectroscopy
- Isotope Tracing
Background:
- Accelerator Mass Spectrometry (AMS) is the standard for 14C quantitation but is costly and has low throughput.
- Laser-based methods offer a potentially more accessible and efficient alternative.
- Two-color cavity ringdown spectroscopy (2C-CRDS) has shown promise for sensitive optical measurements of 14CO2.
Purpose of the Study:
- To demonstrate the feasibility of a mid-IR 2C-CRDS detector for low-level 14C isotope tracing in biological and pharmaceutical applications.
- To assess the sensitivity, accuracy, and throughput of the 2C-CRDS system for 14C quantitation.
- To evaluate the potential of 2C-CRDS as a cost-effective, high-throughput alternative to AMS.
Main Methods:
- Utilized a mid-IR 2C-CRDS system to measure 14CO2 absorption signals from combusted samples.
- Applied the 2C-CRDS 14C sensor to biochemical analysis and pharmaceutical studies.
- Investigated a flow-through 14CO2 sampling scheme to improve measurement throughput.
Main Results:
- Achieved detection sensitivity and quantitation accuracy competitive with AMS, at a few parts per quadrillion 14C/C mole fraction.
- Demonstrated sub-fCi level quantitation of sample 14C activity with a minimum sample size of 3 mg of carbon.
- Current throughput is ~25 min/sample, limited by sample processing, with potential for improvement to a few minutes per sample.
Conclusions:
- The 2C-CRDS 14C sensor is a feasible and competitive technology for low-level isotope tracing.
- The system offers high sensitivity and accuracy, making it suitable for biological, pharmaceutical, and clinical studies.
- Potential for revolutionizing 14C tracing due to improved efficiency, lower cost, and compact size.