Related Experiment Video
Updated: May 24, 2025

Autonomous and Rechargeable Microneurostimulator Endoscopically Implantable into the Submucosa
Published on: September 27, 2018
A Batteryless Implantable System with Adaptive Near-Field Communication to Study Neurogastroenterological Disorders
Abstract:
To study the neurophysiology of the gastric system in the long term, it is essential to develop batteryless and wireless gastric implants with adaptive communication. This paper presents the development of a system that establishes a closed-loop wireless power transfer and bidirectional data transfer between a wearable unit (WU) and a miniature implantable unit (IU) through near-field communication. The neurophysiological activity of the gastric system is digitized, modulated through a load-shift keying (LSK) scheme, and transmitted from the IU to the WU over a 13.56 MHz adaptive inductive link. Besides, electroceutical therapeutic instructions are modulated through an amplitude-shift keying (ASK) scheme and sent over the reverse path for execution. A half-duplex data transmission protocol was developed to manage the sequential IU-WU and WU-IU data transfers and avoid collisions of data packets over the same inductive link. In addition, the uplink and downlink data are encoded through a self-clocking algorithm. The benchtop validation of the developed system demonstrated successful near-field communication at a data transfer rate of 125 kbps with no mismatch in both directions.