Related Experiment Video
Updated: May 24, 2025

Author Spotlight: Development and Evaluation of a Standardized Rat Model for Calvarial Suture-Bony Composite Defects
Published on: May 10, 2024
The Not Fade Away Phenomenon: 2 Cases of Delayed Nonabsorption of Biodegradable Fixation in Craniosynostosis
Aulon Jerliu1, Iulianna C Taritsa2, Jose Foppiani2
1From the Department of Surgery, University of Connecticut School of Medicine, Farmington, CT.
Abstract:
Bioresorbable plating systems have been more recently developed as an alternative to metal osteofixation systems and can be used for craniosynostosis operations. To date, the bioresorbable materials used for osteofixation have been used clinically with excellent safety outcomes and good patient outcomes without major plate-induced complications. In this report, we present 2 cases in which bioresorbable materials failed to degrade following operations for craniosynostosis in a chronic manner mandating removal and revision. We also provide a review of the existing literature regarding bioresorbable fixation and evidence of failure to degrade. Based on the clinical courses and pathological findings, it is likely that material factors, combined with the metabolic resorptive capacity and the patient host immune system, contributed to the failure of resorption. Treatment with the removal of implant debris has proven to be effective in other reports, and both patients ultimately progressed with acceptable clinical outcomes. Evaluating rare complications associated with these systems will help provide evidence to inform future product development and refinement. As the field of craniofacial surgery continues to evolve, the use of new materials and technologies promises to enhance patient outcomes, reduce complications, and expand the scope of surgical interventions.
More Related Videos
07:35Creating Rigidly Stabilized Fractures for Assessing Intramembranous Ossification, Distraction Osteogenesis, or Healing of Critical Sized Defects
Published on: April 11, 2012
04:19Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
Published on: November 8, 2024