Basic verification of myocardial extracellular volume quantification by prototype photon-counting detector computed
Seitaro Oda1, Yoshinori Funama2, Shinichi Kojima3
1Departments of Diagnostic Radiology, Kumamoto University, Kumamoto, Japan.
Journal of Clinical Imaging Science
|March 5, 2025
Summary
Photon-counting detector (PCD) CT accuracy for myocardial extracellular volume (ECV) quantification depends on radiation dose and virtual monoenergetic image (VMI) settings. Optimizing keV levels and dose is crucial for precise ECV measurements.
Area of Science:
- Medical Imaging
- Radiology
- Cardiovascular Imaging
Background:
- Myocardial extracellular volume (ECV) quantification is vital for assessing cardiac diseases.
- Photon-counting detector (PCD) computed tomography (CT) offers advanced spectral imaging capabilities.
- Understanding the impact of radiation dose and spectral settings on ECV accuracy is essential.
Purpose of the Study:
- To evaluate the accuracy of myocardial ECV quantification using a prototype PCD CT.
- To investigate the influence of radiation dose and virtual monoenergetic image (VMI) settings on ECV measurements.
- To determine optimal spectral settings for reliable ECV quantification with PCD CT.
Main Methods:
- A multi-energy CT phantom simulating blood pool and myocardium was utilized.
- Scans were performed with varying tube current-time products (105, 150, 300 mAs) at 120 kVp.
- Virtual monoenergetic images (VMIs) from 50-100 keV were reconstructed to calculate ECV.
Main Results:
- Radiation dose and VMI settings significantly affected ECV quantification accuracy.
- ECV values were overestimated at higher keV levels across all dose settings.
- Higher keV levels showed increased ECV variability, especially at lower radiation doses.
Conclusions:
- Appropriate VMI keV and radiation dose settings are necessary for accurate myocardial ECV quantification with PCD CT.
- The choice of keV impacts both the ECV quantification value and measurement variability.
- Careful optimization of imaging parameters is required to minimize errors in ECV assessment using PCD CT.
Keywords:
Myocardial extracellular volumePhantom studyPhoton-counting detector computed tomographyVirtual monoenergetic imaging