Systematic Hearing Performance Evaluation Process for Adolescents with Cochlear Implantation at Early Ages
Published on: March 24, 2023
324
A novel cochlear implant assessment tool: Audiometric and speech recognition analysis
Fernanda Ferreira Caldas1, Byaka Cagnacci Buzo2, Bruno Sanches Masiero3
1Universidade de Brasília, Faculdade de Ciências da Saúde, Brasilia, DF, Brazil; Instituto Brasiliense de Otorrinolaringologia, Brasilia, DF, Brazil.
Brazilian Journal of Otorhinolaryngology
|March 6, 2025
Summary
CLABOX with direct audio input effectively evaluated cochlear implant (CI) users for pure tone audiometry and speech recognition, showing comparable results to traditional sound booths. This offers a viable alternative for CI audiological assessments.
Area of Science:
- Audiology
- Biomedical Engineering
- Speech-Language Pathology
Background:
- Cochlear implants (CIs) are crucial for hearing restoration in individuals with severe to profound hearing loss.
- Accurate audiological assessments, including pure tone audiometry (PTA) and speech recognition testing, are vital for optimizing CI outcomes.
- Traditional sound booth (SB) evaluations can be time-consuming and resource-intensive.
Purpose of the Study:
- To compare the efficacy of CLABOX with direct audio input (DAI) against conventional sound booth (SB) evaluations for assessing PTA and speech recognition in CI users.
- To determine if CLABOX-DAI can serve as a reliable alternative for audiological testing in the cochlear implant population.
Main Methods:
- Fifty cochlear implant users (33 adults, 17 children) participated in the study.
- Speech recognition tests in noise (fixed and adaptive), Hearing in Noise Test (HINT), Ling test, and PTA were administered in both the SB and CLABOX environments.
- Data were analyzed to compare performance metrics between the two testing setups.
Main Results:
- Speech recognition in noise and PTA thresholds were significantly better when measured in the CLABOX compared to the SB.
- The signal-to-noise ratio (S/N) for HINT with adaptive noise was more favorable in CLABOX (2.14 dB) than in SB (3.42 dB).
- No significant difference was observed for the Ling test categories between the two environments.
Conclusions:
- CLABOX with DAI is an effective tool for evaluating PTA and speech recognition in cochlear implant users.
- The findings suggest CLABOX-DAI can be a valuable and potentially more efficient alternative to traditional sound booths for CI audiological assessments.